Simultaneous non‐invasive assessment of arterial stiffness and haemodynamics – a validation study

View Full Paper
TKTiit KööbiMKMika KähönenTITiina Iivainen

Key Result

Whole-body impedance cardiography-derived pulse wave velocity agreed well with Doppler ultrasound (bias 0.00 m/s) and correlated strongly with changes in PWV (r=0.74; P<0.0001).

Study Design

Type

Observational (n=55)

Structured PICO

Does whole-body impedance cardiography provide reliable measurements of arterial stiffness and hemodynamics compared to Doppler ultrasound and invasive methods in healthy subjects and surgical patients?

P
Population
25 healthy subjects and 30 surgical patients
I
Intervention
Whole-body impedance cardiography (ICGWB) for measuring pulse wave velocity (PWV) and stroke volume index to pulse pressure (SI/PP)
C
Comparator
Doppler ultrasound (for PWV) and thermodilution/direct Fick methods (for SI/PP)
O
Outcome
Agreement, repeatability, and reproducibility of PWV and SI/PP measurementssurrogate

Whole-body impedance cardiography is a reliable and reproducible non-invasive method for evaluating arterial stiffness and hemodynamics.

Main Result

Effect estimate: r=0.74

p-value: p=<0.0001

Abstract

The purpose of the study was to estimate the reliability of whole-body impedance cardiography (ICGWB)-derived pulse wave velocity (PWV) and stroke volume index to pulse pressure (SI/PP) measurements. The repeatability and reproducibility of ICGWB parameters were also determined. Agreement between the impedance and Doppler ultrasound-based PWV measurements was estimated in 25 healthy subjects in two consecutive measurements. Impedance-derived SI/PP (SIICG/PP) estimates were compared with simultaneously measured SI/PP based on thermodilution (SITD/PP) and direct Fick (SIFICK/PP) methods in 30 surgical patients. PWV measured between the aortic arch and popliteal artery using the impedance technique with selective electrode configuration (PWVIS) agreed well with the Doppler ultrasound method (PWVDOPP), the bias (PWVDOPP - PWVIS) and precision (+/- SD of differences) being 0.00 and 0.79 m s-1, respectively. PWV derived from the whole-body and popliteal impedance plethysmograms (PWVICG) overestimated slightly PWVDOPP values. The repeatability value for PWVIS was excellent, being 0.54 m s-1. The reproducibility values for PWVDOPP and PWVIS were very similar (2.17 and 2.42 m s-1, respectively). Changes in PWVIS correlated strongly with changes in PWVDOPP (r=0.74; P<0.0001), indicating that both methods reflected the true physiological variation in PWV. The agreement between SIICG/PP and SITD/PP or SIFICK was almost identical to the agreement between the SITD/PP and SIFICK/PP. In conclusion,whole-body impedance cardiography provides handy and reliable means of evaluating arterial stiffness on the basis of PWV and SI/PP simultaneously with conventional haemodynamic parameters. The method is highly repeatable and reproducible.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Kööbi et al. (2003) conducted an observational in Healthy subjects and surgical patients (n=55). Whole-body impedance cardiography (ICGWB) vs. Doppler ultrasound, thermodilution, and direct Fick methods was evaluated on Agreement between impedance and Doppler ultrasound-based PWV measurements (r=0.74, p=<0.0001). Whole-body impedance cardiography-derived pulse wave velocity agreed well with Doppler ultrasound (bias 0.00 m/s) and correlated strongly with changes in PWV (r=0.74; P<0.0001).