Degradation of biodegradable orthopedic screws does (not) elevate tissue heating under Magnetic Resonance Imaging: Implications for next generation implants

View Full Paper
MBMostafa BerangiHWHelmar WaicziesNSNandita Saha

Key Points

Key points are not available for this paper at this time.

Abstract

by up to 19.1% but redistributed heat more uniformly across the implant, lowering peak temperatures compared to intact screws. In the human voxel model, fractured screws remained below the IEC safety limits under maximum permissible RF exposure with a maximum temperature elevation of 38.9 °C, whereas non-fractured screws exceeded 40 °C. Our results demonstrate that Mg screw degradation generally mitigates MRI heating risks. In conclusion, engineered degradation features may enhance MRI safety, providing design guidance for next-generation biodegradable orthopedic implants.