November 1, 1991Pacing and Clinical Electrophysiology23 citations

Accuracy of Rhythm Classification Using a Data Log System in Implantable Cardioverter Defibrillators

View Full Paper
PWPaul J. WangNMNICHOLAS J. MANDALAKASCCChristopher A. Clyne

Structured PICO

Does the use of a data log system improve the accuracy of rhythm classification in patients with implantable cardioverter defibrillators compared to symptoms alone?

P
Population
20 patients with 32 separate tachyarrhythmia episodes
I
Intervention
Rhythm classification using data log information (Telectronics ATP 4210) alone or combined with symptoms
C
Comparator
Rhythm classification based on symptoms alone
O
Outcome
Sensitivity, specificity, and predictive accuracy of rhythm classificationsurrogate

Data log information from ICDs significantly improves the accuracy of tachyarrhythmia rhythm classification compared to relying on patient symptoms alone.

Abstract

Because the presence or absence of symptoms alone may be insufficient to correctly diagnose the rhythm for which implantable cardioverter defibrillator therapy is delivered, we hypothesized that the addition of data log information available in Telectronics ATP 4210 may improve the accuracy of rhythm classification. With this system the recorded ventricular electrogram cycle length is reported on a beat-to-beat basis immediately before, during, and after the tachyarrhythmia is detected. Using this information recorded from the data log in 32 separate tachyarrhythmia episodes in 20 patients, we compared the sensitivity, specificity, and predictive accuracy of rhythm classification on the basis of symptoms alone, data log alone, and data log combined with symptoms. While classification based on symptoms alone is highly specific (10/10 episodes), it is insensitive and has an overall predictive accuracy of 53%. By contrast, data log is sensitive (90%) and specific (91%) with better predictive accuracy (94%) than symptoms alone (P = 0.002). The addition of symptoms to information on beat-to-beat cycle length from data log resulted in a slight increase in predictive accuracy.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Wang et al. (1991) studied this question.

synapsesocial.com/papers/6a1ccc69732133fe9885e63ehttps://doi.org/10.1111/j.1540-8159.1991.tb02789.x
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1The automatic implantable cardioverter-defibrillator. Clinical experience, complications, and follow-up in 25 patients1988 · 42 citations
  2. 2Clinical experience, complications, and survival in 70 patients with the automatic implantable cardioverter/defibrillator.1985 · 453 citations
  3. 3Mortality in Patients with Implanted Automatic Defibrillators1983 · 313 citations
  4. 4Implantable Defibrillation: Eight Years Clinical Experience1988 · 143 citations
  5. 5Termination of Malignant Ventricular Arrhythmias with an Implanted Automatic Defibrillator in Human Beings1980 · 1,371 citations