XPS: binding energy calibration of electron spectrometers 5?re-evaluation of the reference energies

View Full Paper
MSM. P. SeahIGIan S. GilmoreGBG. Beamson

Key Points

Key points are not available for this paper at this time.

Abstract

The binding energies of the calibration peaks for x-ray photoelectron spectroscopy—Cu 2p3/2, Ag 3d5/2 and Au 4f7/2—have been reassessed based on the traceable data recorded in 1984 using unmonochromated x-rays and an analyser resolution of 0.3 eV. The changes in those energies, for different x-ray sources and analyser resolutions, have been calculated and the results compared with further data. This includes work with monochromatic Al x-rays recorded at high energy resolution, allowing the binding energies to be referred to a new zero value set at the Fermi edge measured for Ag. A consistent set of data is presented for the calibration and assessment of photoelectron spectrometers with energy resolutions in the range 0.2–1.5 eV, when used with unmonochromated Al or Mg x-rays or monochromated Al x-rays. © 1998 John Wiley & Sons, Ltd.