January 25, 2008International Journal of Sports Medicine129 citations

Validity of the Polar S810 to Measure R-R Intervals in Children

View Full Paper
FGFrançois-Xavier GamelinGBGeorges BaquetSBSerge Berthoin

Structured PICO

Does the Polar S810 heart rate monitor accurately measure R-R intervals and HRV compared to ECG in children?

P
Population
12 children, mean age 9.6 +/- 0.9 years
I
Intervention
Polar S810 heart rate monitor (HRM) in a supine position
C
Comparator
Electrocardiogram (ECG) in a supine position
O
Outcome
R-R intervals and heart rate variability (HRV) parameterssurrogate

The Polar S810 heart rate monitor provides valid measurements of R-R intervals and heart rate variability in children in a supine position compared to standard ECG.

Abstract

Intervals between two consecutive cardiac beats (R-R intervals) and the subsequent analysis of heart rate variability (HRV) obtained simultaneously from the Polar S810 heart rate monitor (HRM) and an electrocardiogram (ECG) in a supine position were compared in twelve children (age 9.6 +/- 0.9 years) before and after protocol correction. R-R intervals were significantly different between the ECG and the HRM uncorrected and corrected signal (p 0.05) and well correlated (r > 0.99, p < 0.05), except SD2 (p < 0.05, ES = 0.000; r = 0.99). These data support the validity of the Polar S810 HRM to measure R-R intervals and make the subsequent HRV analysis in a supine position in children.