Ambulatory blood pressure monitoring by a novel cuffless device: a pilot study

View Full Paper
FSFilippo ScaliseDMDavide MargonatoASAndrea Sole

Key Result

A novel cuffless blood pressure device estimated 24-hour mean SBP and DBP higher than a standard oscillometric device (131.1/80.2 vs 125.4/75 mmHg), with moderate correlation (SBP r=0.58, DBP r=0.65).

Study Design

Type

Observational (n=84)

Structured PICO

Does a novel cuffless blood pressure device accurately measure 24-hour ambulatory blood pressure compared to a standard oscillatory cuff-based device in consecutive patients?

P
Population
84 consecutive patients undergoing 24-hour ambulatory blood pressure monitoring.
E
Exposure
Cuffless blood pressure (CL-BP) device for 24-hour ambulatory blood pressure monitoring
C
Comparator
Standard oscillatory cuff-occlusion-based (C-BP) device
O
Outcome
Accuracy and correlation of 24-hour mean systolic blood pressure, diastolic blood pressure, and heart ratesurrogate

A novel cuffless blood pressure device showed moderate correlation with standard oscillometric devices for 24-hour ambulatory monitoring, with better accuracy in male and overweight patients, but estimated mean pressures differently.

Main Result

Effect estimate: r = 0.58 (SBP), r = 0.65 (DBP)

Absolute Event Rate: 131.1% vs 125.4%

Abstract

Purpose Cuffless blood pressure (CL-BP) measurements are believed to be a potentially alternative to cuff-occlusion-based (C-BP) measurement. A new cuffless device was developed for ambulatory BP monitoring. We assessed the accuracy of a new CL-BP device compared to a standard oscillatory C-BP device over the 24 h.Materials and methods Eighty-four consecutive patients were included in the study. BP was measured simultaneously by the CL-BP device and by a C-BP device over the 24 h. Calculations included 24 h mean systolic (S) BP, the mean diastolic (D) BP and the heart rate (HR). Correlations between the CL-BP and C-BP measurements were sought using Pearson’s correlation coefficients and Bland-Altman plots.Results Using the C-BP device, the 24 h SBP value for the cohort was 125.4 ± 10.9 mmHg (mean ± SD); the corresponding DBP value being 75 ± 8.3 mmHg. Mean SBP/DBP were higher with the CL-BP device, i.e. 131.1 ± 15.9/80.2 ± 9.7 mmHg . The correlation coefficients between the two sets of values were significant (SBP: r = 0.58, DBP: r = 0.65). Better correlations for SBP and DBP were found 1) in patients with BMI > 25 (SBP: r = 0.65, DBP: r = 0.70) compared to those with BMI <25 and 2) in males compared to females (SBP: r = 0.71, DBP: r = 0.77).Conclusions In our patients a CL-BP device estimated 24 h mean SBP and DBP differently from the classical oscillometric device, with a moderate correlation. CL-BP measurements were most accurate on male and overweight subjects.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Scalise et al. (2020) reported an observational. Cuffless blood pressure (CL-BP) device vs. Standard oscillatory cuff-occlusion-based (C-BP) device was evaluated on 24 h mean systolic (S) BP and diastolic (D) BP (r = 0.58 (SBP), r = 0.65 (DBP)). A novel cuffless blood pressure device estimated 24-hour mean SBP and DBP higher than a standard oscillometric device (131.1/80.2 vs 125.4/75 mmHg), with moderate correlation (SBP r=0.58, DBP r=0.65).