Evaluation of a New Ambulatory Spirometer for Measuring Forced Expiratory Volume in One Second and Peak Expiratory Flow Rate

View Full Paper
DJDP JohnsMAMichael J. AbramsonGBGlenn Bowes

Key Points

Key points are not available for this paper at this time.

Abstract

A new pocket-sized ambulatory spirometer (meter) that measures FEV1 and PEFR was evaluated. The meter determines flow rate (V) from the differential pressure (P1-P2) across a sharp-edged orifice (V alpha square root ofP1-P2). Ten meters were evaluated for accuracy using a computerized syringe to deliver the 24 ATS-recommended waveforms and nine scaled versions of a single waveform (PW#24). Within-meter reproducibility of FEV1 and PEFR was evaluated in two meters by passing five versions of waveform PW#24 through each meter 10 times. Reliability in the ambulatory setting was assessed in six meters on several occasions over a 10-week period using five versions of waveform PW#24. Results show that the 10 meters conform to the ATS accuracy specifications for PEFR with one or less errors and marginally outside these limits for FEV1 with four errors. For the nine versions of PW#24, the 95% confidence intervals indicate that the meter is accurate to within +/- 5.5% or +/- 15 L/min for PEFR and +/- 3.5% or +/- 0.12 L for FEV1. The mean within-meter coefficient of variation was 1.24% for FEV1 and 0.35% for PEFR. There was no significant change in meter accuracy or performance over the 10-wk reliability study. We conclude that the meter is suitable for use as an ambulatory spirometer for measuring FEV1 and PEFR.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Johns et al. (1993) studied this question.

synapsesocial.com/papers/6a1f5ca1da07783dec235894https://doi.org/10.1164/ajrccm/147.5.1245
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Evaluation of Accuracy and Reproducibility of Peak Flowmeters at 1,400 m1992 · 65 citations
  2. 2Comparison of PEFR and FEV1 in Patients with Varying Degrees of Airway Obstruction1989 · 70 citations
  3. 3Measurement of FEV 1 and FVC1988 · 15 citations
  4. 4Effort and the Forced Expiratory Volume in One Second (FEV1)1970 · 7 citations
  5. 5Maximum Forced Expiratory Flow Rate as a Measure of Ventilatory Capacity1959 · 469 citations