August 25, 2005Analytical Chemistry264 citations

Reusable Impedimetric Aptasensor

View Full Paper
ARAbd‐Elgawad RadiJSJosep Lluís Acero SánchezEBEva Baldrich

Key Points

Key points are not available for this paper at this time.

Abstract

A novel impedimetric aptasensor using a mixed self-assembled monolayer composed of thiol-modified thrombin binding aptamer and 2-mercaptoethanol on a gold electrode is reported. The changes of interfacial features of the electrode were probed in the presence of the reversible redox couple, Fe(CN)6(3-/4-), using impedance measurements. The electrode surface was partially blocked due to the self-assembly of aptamer or the formation of the aptamer-thrombin complex, resulting in an increase of the interfacial electron-transfer resistance detected by electrochemical impedance spectroscopy or cyclic voltammetry. The aptasensor was regenerated by breaking the complex formed between the aptamer and thrombin using 2.0 M NaCl solution, and the immobilized aptamer subsequently was used for repeated detection of thrombin. The aptamer-functionalized electrode showed a linear response of the charge-transfer resistance to the increase of thrombin concentration in the range of 5.0-35.0 nM and the thrombin was easily detectable to a concentration of 2.0 nM.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Radi et al. (2005) studied this question.