Respiration as a reliable physiological sensor for controlling cardiac pacing rate.

PRPaolo RossiGPGianni PlicchiGCG Canducci

Structured PICO

Does respiration-rate-controlled cardiac pacing improve exercise capacity and physiological parameters in patients dependent on pacemakers?

P
Population
73 patients for correlation study (19 with normal lung function, 4 with restrictive pulmonary disease, and 50 with obstructive airways disease); 11 patients with ventricular inhibited pacemakers tested with an external computerized system; 2 patients implanted with a prototype respiration-dependent pacemaker.
I
Intervention
Variable cardiac pacing rate controlled by respiration rate (using an external computerized programmable system or implanted prototype) during exercise stress tests.
C
Comparator
Standard pacing without respiration-rate control (implied intra-patient comparison during exercise tests).
O
Outcome
Maximum values of oxygen uptake, minute ventilation, and work time during exercise stress tests.surrogate

Respiration rate serves as a valid and stable physiological variable for rate-responsive pacing, improving exercise capacity in pacemaker-dependent patients.

Abstract

A study was carried out to determine whether variations in the respiration rate during physical exercise could be used as a physiological variable in controlling the rate of an implanted pacemaker. The relation between respiration rate and heart rate was significantly correlated in 73 patients (19 with normal lung function, four with restrictive pulmonary disease, and 50 with obstructive airways disease) during repeated calibrated ergometric tests; no significant differences were found between the subgroups. An external computerised programmable system with algorithm control activated by a radio frequency system was used to vary the cardiac stimulation rate in relation to respiration rate in 11 patients implanted with ventricular inhibited pacemakers. In addition, a prototype programmable pacemaker dependent on respiration rate was implanted in two patients. Maximum values of oxygen uptake, minute ventilation, and work time were increased during the exercise stress tests when the variable cardiac pacing rate was used. Thus respiration rate appears to be a valid and stable physiological variable for controlling the cardiac stimulation rate in order to improve cardiac output in patients dependent on pacemakers.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Rossi et al. (1984) studied this question.

synapsesocial.com/papers/6a2086ebff6a0d6103f89a30https://doi.org/10.1136/hrt.51.1.7
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Characterization of the Circulatory Response to Maximal Upright Exercise in Normal Subjects and Patients with Heart Disease1967 · 161 citations
  2. 2Ventilatory responses to the metabolic acidosis of treadmill and cycle ergometry1976 · 110 citations
  3. 3Exercise testing of patients with coronary heart disease. Principles and normal standards for evaluation.1971 · 596 citations
  4. 4Die Methoden der röntgenologischen Herzvolumenbestimmung und ihre Fehlerbreite*1964 · 13 citations
  5. 5An implantable synchronous pacemaker for the long term correction of complete heart block1963 · 133 citations