Percutaneous aortic valve replacement: resection before implantation☆

View Full Paper
RQRené QuadenTATim AttmannABAndreas Boening

Key Result

High-pressure water stream resection of human calcified aortic valves took a mean of 6.0 minutes at 150 bar, demonstrating feasibility for endoscopic cutting.

Structured PICO

Is high-pressure water stream resection feasible and safe for endovascular resection of calcified aortic valves in an in vitro model?

P
Population
In vitro study evaluating the resection of 17 human calcified aortic valves and 20 healthy porcine aortic valves using a high-pressure water stream.
I
Intervention
High-pressure water stream resection (water jet) at 60 to 150 bar
O
Outcome
Feasibility of resection (time taken, size of cut leaflets) and destruction to surrounding anatomysafety

High-pressure water stream resection is a feasible auxiliary method for cutting calcified aortic valves endoscopically, though a debris-capturing device is required before clinical use.

Limitations

  • An additional aortic valve resection chamber designed for capturing all debris has to be established before clinical application.
  • Requires an additional aortic valve resection chamber to capture debris before clinical application

Abstract

OBJECTIVE: After transluminal endovascular implantation of a new valved stent, the aim of this study was to evaluate the feasibility of using a high-pressure water stream to endovascularly resect human calcified aortic valves. METHODS: First, human calcified aortic valves were excised and then resected in vitro to determine optimal water jet parameters. Second, healthy porcine aortic valves were ablated in vitro to evaluate possible destruction to the surrounding anatomy. Third, resection was performed endoluminally by introducing microsystemic tools into the descending aorta, passing them through the arch and ascending aorta to the aortic valve in an in vitro porcine model. Macro- and micropathology of specimens were analyzed. RESULTS: First, resection of human calcified valves took a mean of 6.0+/-2.4min per three leaflets at 150bar (n=17). The maximum size of the cut leaflets was 7.1+/-1.7mm. Second, resection of healthy porcine aortic valves at 60bar took 2.3+/-0.3min per three leaflets (n=10). Only the aortic annulus was moderately affected in six cases. Third, endoluminal resection via the descending aorta took 12.2+/-0.8min per three leaflets at 60bar (n=10). The aortic wall was affected in four cases, the aortic annulus and the coronary ostia only once. Microscopic analysis also revealed superficial lesions with a maximum lesion depth of 1200microm in one case, and an average of 580+/-145microm in subsequent lesions. The mitral valve and the left ventricular outflow tract were not affected. CONCLUSIONS: Percutaneous resection of heart valves is emerging as a promising auxiliary method for the resection of calcified aortic heart valves because they can be cut endoscopically. Nonetheless, before this resection tool can be clinically applied by surgeons to perform a true percutaneous valve replacement, an additional aortic valve resection chamber (already at the prototype stage) designed for capturing all debris, has to be established.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Quaden et al. (2005) studied Calcified aortic valves (n=37). High-pressure water stream resection was evaluated on Resection time and tissue damage. High-pressure water stream resection of human calcified aortic valves took a mean of 6.0 minutes at 150 bar, demonstrating feasibility for endoscopic cutting.

synapsesocial.com/papers/6a20a595e033bce76a911f28https://doi.org/10.1016/j.ejcts.2005.01.048
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Steps Toward Percutaneous Aortic Valve Replacement2002 · 154 citations
  2. 2Transluminal implantation of artificial heart valves. Description of a new expandable aortic valve and initial results with implantation by catheter technique in closed chest pigs1992 · 492 citations
  3. 3Transcatheter Implantation of a Bovine Valve in Pulmonary Position2000 · 320 citations
  4. 4Development and initial experimental evaluation of a prosthetic aortic valve for transcatheter placement. Work in progress.1992 · 87 citations
  5. 5Do valved stents compromise coronary flow?☆2004 · 45 citations