PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
February 14, 2023Sensors32 citationsOpen Access

Signal Quality Analysis for Long-Term ECG Monitoring Using a Health Patch in Cardiac Patients

View Full Paper
ICIsrael Campero-JuradoILIlde LoratoJMJohn Morales

Key Result

Long-term ECG monitoring using a single-lead health patch in 6 cardiac patients showed no statistically significant degradation in average daily signal quality over 5 days.

Study Design

Type

Observational (n=6)

Structured PICO

Does a single-lead ECG patch maintain adequate signal quality over 5 days of monitoring in cardiac patients?

P
Population
6 cardiac patients monitored with a single-lead ECG patch for 5 days.
E
Exposure
Single-lead ECG patch featuring self-adhesive dry electrode technology for 5 days
O
Outcome
Signal quality degradation over timesurrogate

Single-lead ECG patches provide stable signal quality over 5 days of continuous monitoring in cardiac patients, supporting their use for long-term ambulatory arrhythmia detection.

Abstract

Cardiovascular diseases (CVD) represent a serious health problem worldwide, of which atrial fibrillation (AF) is one of the most common conditions. Early and timely diagnosis of CVD is essential for successful treatment. When implemented in the healthcare system this can ease the existing socio-economic burden on health institutions and government. Therefore, developing technologies and tools to diagnose CVD in a timely way and detect AF is an important research topic. ECG monitoring patches allowing ambulatory patient monitoring over several days represent a novel technology, while we witness a significant proliferation of ECG monitoring patches on the market and in the research labs, their performance over a long period of time is not fully characterized. This paper analyzes the signal quality of ECG signals obtained using a single-lead ECG patch featuring self-adhesive dry electrode technology collected from six cardiac patients for 5 days. In particular, we provide insights into signal quality degradation over time, while changes in the average ECG quality per day were present, these changes were not statistically significant. It was observed that the quality was higher during the nights, confirming the link with motion artifacts. These results can improve CVD diagnosis and AF detection in real-world scenarios.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Campero-Jurado et al. (2023) conducted an observational in Cardiac patients (n=6). Single-lead ECG patch featuring self-adhesive dry electrode technology was evaluated on Signal quality degradation over time. Long-term ECG monitoring using a single-lead health patch in 6 cardiac patients showed no statistically significant degradation in average daily signal quality over 5 days.

synapsesocial.com/papers/6a20da91ff90007ec6ecb14chttps://doi.org/10.3390/s23042130
Ask AI
Helpful
Bookmark
Share
View Full Paper