Dosimetric Impact of Systematic Hounsfield Units Shifts on Prostate VMAT Treatment Planning

View Full Paper
HIHattal Imane

Key Points

  • This study evaluates how systematic shifts in Hounsfield Units affect dosimetric accuracy in VMAT for prostate cancer.
  • Established a reference HU-RED calibration curve using CIRS Model 062M phantom with Philips CT scanner.
  • Modified curves with systematic shifts of +50 HU and +200 HU were applied to ten prostate cancer treatment plans.
  • Compared dosimetric parameters using Wilcoxon signed-rank test.
  • +50 HU shift showed minimal dosimetric changes (<0.5%), p>0.05.
  • +200 HU shift resulted in significant reductions: PTV Dmax decreased by 1.55% (p=0.037) and Dmean by 1.26% (p=0.043).
  • Bladder sensitivity showed V25Gy variations up to 38.15% (p=0.005), while rectal parameters remained stable, p>0.05.

Abstract

Background: Accurate conversion of computed tomography (CT) Hounsfield Units (HU) into relative electron density (RED) is fundamental for precise dose calculation in external beam radiotherapy. Variations in the HU-RED calibration curve, arising from scanner drift, acquisition parameters, or reconstruction algorithms, may introduce dosimetric uncertainties that could affect treatment outcomes. Purpose: This study aimed to systematically evaluate the dosimetric impact of HU calibration variations on volumetric modulated arc therapy (VMAT) treatment plans for prostate cancer, focusing on target coverage and organs at risk (OARs) sparing. Methods: A reference HU-RED calibration curve was established using a CIRS Model 062M electron density phantom scanned with a Philips Brilliance Big Bore CT scanner. Two modified curves were generated by applying systematic shifts of +50 HU and +200 HU. Ten patients with localized prostate cancer, treated with VMAT (78 Gy in 39 fractions), were retrospectively selected. Three treatment plans were generated per patient using each calibration curve. Dosimetric parameters including Dmax, Dmin, Dmean, conformity index (CI), and homogeneity index (HI) for the planning target volume (PTV), along with dose-volume metrics for bladder, rectum, and femoral heads, were compared using the Wilcoxon signed-rank test. Results: The +50 HU shift produced minimal dosimetric changes (0.05). In contrast, the +200 HU shift resulted in statistically significant reductions in PTV Dmax (1.55%, p=0.037) and Dmean (1.26%, p=0.043), along with significant deterioration in CI (6.89% increase, p=0.008). The bladder showed the highest sensitivity with V25Gy variations up to 38.15% (p=0.005), while rectal parameters remained stable across all curves (p>0.05). Conclusion: HU variations within +50 HU maintain dosimetric accuracy within clinically acceptable thresholds. However, larger calibration errors (+200 HU) can produce significant dosimetric deviations, particularly for soft-tissue OARs. These findings emphasize the importance of rigorous CT calibration protocols and regular quality assurance to ensure treatment planning accuracy in prostate radiotherapy.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Hattal Imane (2026) studied this question.

synapsesocial.com/papers/6a2117bfd499ed480b170932https://doi.org/10.5281/zenodo.20502908
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Dosimetric Impact of Systematic Hounsfield Units Shifts on Prostate VMAT Treatment Planning2026
  2. 2Evaluation of the impact of anatomical region-specific HU-RED conversion curves on VMAT treatment planning for head and neck cancer2026
  3. 3HU-to-density calibration curves using diagnostic CT for radiotherapy treatment planning: a feasibility study2026 · 1 citations
  4. 4Evaluation of the accuracy of cone-beam CT–based dose calculation for target volumes and organs at risk in left-sided breast cancer radiotherapy2026
  5. 5The effect of interfractional variation on delivered dose with ultrahypofractionated pencil beam scanning proton therapy for localized prostate cancer2025 · 1 citations