Evaluation of a transparent blood analog fluid: Aqueous Xanthan gum/glycerin

View Full Paper
KBK.A. BrookshierJTJohn M. Tarbell

Key Result

Peak wall shear rates using aqueous Xanthan gum/glycerin solutions were not statistically different from blood in straight and atherogenic curved artery models (p > 0.05).

Structured PICO

P
Population
In vitro straight and atherogenic curved artery models
I
Intervention
Aqueous Xanthan gum/glycerin solution (blood analog fluid)
C
Comparator
Blood
O
Outcome
Peak wall shear rates in pulsatile flowsurrogate

Aqueous Xanthan gum/glycerin solutions provide a transparent, inexpensive, and accurate blood analog fluid for in vitro hemodynamic studies of large arteries.

Main Result

p-value: p=> 0.05

Abstract

Aqueous solutions of Xanthan gum and glycerin were developed that match both the viscous and elastic components of the 2 Hz complex viscosity of blood over a broad range of shear rates (1-1000 s-1) and hematocrits (20%-80%). Measurements of wall shear rate in pulsatile flow were conducted in straight and atherogenic curved artery models using both the blood analog fluid and blood to compare their flow behavior. Peak wall shear rates determined in these models under nearly identical flow conditions with the two fluids were not statistically different (p > 0.05). These results indicate that aqueous Xanthan gum/glycerin solutions provide a good blood analog fluid for flow conditions characteristic of large arteries. Because the fluid is transparent and made from readily available and fairly inexpensive materials, it may be generally useful for in vitro hemodynamic studies, particularly those employing optical techniques.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Brookshier et al. (1993) studied this question. Aqueous Xanthan gum/glycerin solution vs. Blood was evaluated on Peak wall shear rates in pulsatile flow in straight and atherogenic curved artery models (p=> 0.05). Peak wall shear rates using aqueous Xanthan gum/glycerin solutions were not statistically different from blood in straight and atherogenic curved artery models (p > 0.05).