PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
July 7, 2026Journal of Magnetic Resonance Imaging0 citations

RF Noise Artifacts From a High Frequency Spinal Cord Stimulation System: A Retrospective Imaging Cohort With Illustrative Cases

View Full Paper
ESEric G. StinsonAHAbby HubbardJKJenna E. Kleinow

Key Result

A specific model of spinal cord stimulator was associated with RF noise artifacts in ~50% of MR exams, compared to 1%-8% of exams in patients with other models from the same manufacturer.

Key Points

  • This research aims to identify the origin of RF noise artifacts associated with specific spinal cord stimulators during MR imaging.
  • Retrospective review of MR exams (n=173) performed on patients (n=122) with implanted spinal cord stimulators over two years.
  • Assessment of artifacts related to specific device models of spinal cord stimulators from the same manufacturer.
  • Comparison of the prevalence of artifacts between different stimulator models.
  • RF noise artifacts were observed in ~50% of MR exams for a specific spinal cord stimulator model.
  • In other models from the same manufacturer, RF noise artifacts were found in only ~1%-8% of exams.
  • Awareness of the artifact can lead to reduced service costs and scheduling delays for MR imaging.

Study Design

Type

Cohort (n=122)

Structured PICO

Does a specific model of high frequency spinal cord stimulator cause more RF noise artifacts on MRI compared to other models?

P
Population
122 patients with implanted spinal cord stimulators from a single manufacturer who underwent 173 MR exams over a two-year period.
E
Exposure
Specific model of high frequency spinal cord stimulator
C
Comparator
Other models of spinal cord stimulators from the same manufacturer
O
Outcome
Presence of RF noise artifacts on MR examssurrogate

A specific model of high frequency spinal cord stimulator is associated with a high rate (~50%) of RF noise artifacts on MRI, which can lead to service costs and scheduling delays.

Abstract

ABSTRACT Unexpected RF noise artifact “followed” a patient with an implanted spinal cord stimulator to multiple MR systems. Subsequent testing and clinical scanning showed no artifact, implying that the artifact originated from the implanted device. MR exams ( n = 173) performed over a two‐year period in patients ( n = 122) who had implanted spinal cord stimulators from the same manufacturer were reviewed for artifacts and device models. RF noise artifacts were observed in ~50% of exams in patients with a specific model of spinal cord stimulator, but only ~1%–8% of exams in other models from the same manufacturer. We hope that by describing this artifact, health care professionals can avoid service costs and scheduling delays.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Stinson et al. (2026) conducted a cohort in Implanted spinal cord stimulators (n=122). Specific model of spinal cord stimulator vs. Other models from the same manufacturer was evaluated on RF noise artifacts during MR exams. A specific model of spinal cord stimulator was associated with RF noise artifacts in ~50% of MR exams, compared to 1%-8% of exams in patients with other models from the same manufacturer.

synapsesocial.com/papers/6a4c9711331bc25c9e5f4207https://doi.org/10.1002/jmri.70394
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Safety and Utility of Spinal Magnetic Resonance Imaging in Patients with High-Frequency Spinal Cord Stimulators: A Prospective Single-Centre Study2019 · 13 citations
  2. 2The Rate of Magnetic Resonance Imaging in Patients With Spinal Cord Stimulation2015 · 63 citations
  3. 3Magnetic Resonance Imaging Exploration of the Human Brain During 10 kHz Spinal Cord Stimulation for Failed Back Surgery Syndrome: A Resting State Functional Magnetic Resonance Imaging Study2019 · 34 citations
  4. 4MRI-Compatible Spinal Cord Stimulator Device and Related Changes in Patient Safety and Imaging Artifacts2014 · 20 citations
  5. 5MRI Conditionality in Patients with Spinal Cord Stimulation Devices2016 · 27 citations