August 18, 2005European Journal of Cardio-Thoracic Surgery27 citationsOpen Access

Echocardiographic findings using tissue velocity imaging following passive containment surgery with the Acorn CorCap™ cardiac support device

View Full Paper
AOArne OlssonFBFredrik BredinAFAnders Franco‐Cereceda

Key Result

Passive containment surgery using the CorCap Cardiac Support Device decreased left ventricular size and improved functional status, but was associated with right ventricular dysfunction.

Study Design

Type

Observational (n=12)

PICO

P
Population
12 heart failure patients with dilated cardiomyopathy who received the CorCap Cardiac Support Device during cardiac surgery.
I
Intervention / Comparator
CorCap Cardiac Support Device
O
Primary Outcome
Echocardiographic evaluation of cardiac dimensions and function

Limitations

  • The functional significance of the observed right ventricular dysfunction is unclear

Abstract

OBJECTIVE: To echocardiographically evaluate the effects of passive containment surgery using the CorCap Cardiac Support Device in heart failure patients with dilated cardiomyopathy. METHODS: Twelve patients with dilated cardiomyopathy subjected to cardiac surgery received the Cardiac Support Device. Patients with ischemic cardiomyopathy (n=5) underwent coronary artery bypass surgery receiving 1-3 bypass grafts. In the idiopathic cardiomyopathy group (n=7), mitral valve annuloplasty was performed in five patients while two patients received the Cardiac Support Device only. RESULTS: Following surgery there was a gradual, sustained improvement in cardiac dimensions (decreased left ventricular end-diastolic diameter and left ventricular end-systolic diameter) combined with an increase in functional status (6-min walk and NYHA class). Concomitantly there was a marked decrease in right ventricular function (decrease in tricuspid annular systolic and diastolic velocities) while the left ventricular function (mitral annular systolic and diastolic velocities) and output (ejection fraction, stroke volume) remained unchanged. CONCLUSIONS: Addition of the Cardiac Support Device to conventional cardiac surgery improves patient status and decreases left ventricular size in heart failure patients with dilated cardiomyopathy. The positive effect on left ventricular dimensions is not accompanied by any improvement in cardiac output but rather right ventricular dysfunction, although the functional significance of this is unclear.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Olsson et al. (2005) conducted an observational in Heart failure with dilated cardiomyopathy (n=12). CorCap Cardiac Support Device was evaluated on Echocardiographic evaluation of cardiac dimensions and function. Passive containment surgery using the CorCap Cardiac Support Device decreased left ventricular size and improved functional status, but was associated with right ventricular dysfunction.

synapsesocial.com/papers/6a6e13fa5d37378ac1dbca40https://doi.org/10.1016/j.ejcts.2005.05.027
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Reverse Remodeling of the Failing Ventricle: Surgical Intervention With the Acorn Cardiac Support Device2004 · 18 citations
  2. 2Changes in natriuretic peptides following passive containment surgery in heart failure patients with dilated cardiomyopathy2008 · 3 citations
  3. 3Passive Containment and Reverse Remodeling by a Novel Textile Cardiac Support Device2001 · 105 citations
  4. 4Reversed remodelling in dilated cardiomyopathy by passive containment surgery is associated with decreased circulating levels of endothelin-12006 · 3 citations
  5. 5Acute Ventricular Reduction with the Acorn Cardiac Support Device: Effect on Progressive Left Ventricular Dysfunction and Dilation in Dogs with Chronic Heart Failure2001 · 25 citations