Significance of Supraventricular Tachyarrhythmias in Patients with Implanted Pacing Cardioverter Defibrillators

View Full Paper
CSClaus SchmittKarlsruhe Institute of TechnologyMMMaría Jesús Miranda MonteroSpanish Clinical Research Network
Juraj Melichercik
Juraj MelichercikHeidelberg University

Structured PICO

What is the incidence of inappropriate ICD activation due to supraventricular tachyarrhythmias in patients with implanted cardioverter defibrillators?

P
Population
86 patients treated with an implantable cardioverter defibrillator (ICD) because of sustained ventricular tachycardia (VT) or ventricular fibrillation (VF).
I
Intervention
Implantable cardioverter defibrillator (ICD) (epicardial system in 27 patients, transvenous system with subcutaneous patch in 59 patients)
O
Outcome
Inappropriate activations of the ICD due to supraventricular tachycardiasafety

Inappropriate activation of ICD therapy triggered by intermittent supraventricular tachyarrhythmias is common, highlighting the need for improved detection algorithms.

Abstract

Eighty-six patients were treated with an implantable cardioverter defibrillator (ICD) because of sustained ventricular tachycardia (VT) or ventricular fibrillation (VF). In 27 patients an epicardial system was used, in 59 patients a transvenous system with a subcutaneous patch electrode was implanted. During a mean follow-up time of 17 +/- 9 months, inappropriate activations of the ICD due to supraventricular tachycardia were documented by Holter monitoring in 14 patients (16%). In 8 patients paroxysmal atrial fibrillation (AF), in 2 patients chronic AF, in 1 patient atrial flutter, and in 3 patients sinus tachycardia triggered antitachycardia pacing functions (12 patients) or internal defibrillation (2 patients). In 3 patients (5%) VT was induced by inappropriate antitachycardia pacing. In an additional 18 patients (21%) inappropriate activation of antitachycardia functions due to atrial tachyarrhythmias were suspected based on telemetry readouts or the patient's history. Inappropriate activation of ICD therapy triggered by intermittent supraventricular tachyarrhythmias is common. Further improvements of detection algorithms for supraventricular tachycardia are required in future device generations.