May 1, 1994Catheterization and Cardiovascular Diagnosis61 citations

Bedside monitoring of heparin therapy: comparison of activated clotting time to activated partial thromboplastin time

View Full Paper
JRJonathan ReinerKCKarin S. CoyneCLConor F. Lundergan

Structured PICO

Does bedside aPTT monitoring correlate better with laboratory aPTT and heparin levels compared to ACT in patients receiving intravenous heparin?

P
Population
100 patients receiving intravenous heparin
I
Intervention
Bedside determination of activated partial thromboplastin time (aPTT) using a portable device and activated clotting time (ACT)
C
Comparator
Hospital laboratory aPTT and heparin levels
O
Outcome
Correlation between bedside aPTT, laboratory aPTT, ACT, and heparin levelssurrogate

Bedside determination of aPTT correlates excellently with laboratory aPTT and is superior to ACT for monitoring intravenous heparin therapy, particularly when maintenance of therapeutic anticoagulation is critical.

Abstract

Heparin anticoagulation is utilized during and after interventional cardiac catheterization procedures to reduce the risk of acute thrombotic coronary artery occlusion. The short half-life of heparin, the importance of maintaining therapeutic anticoagulation, and the time delay inherent in the processing and retrieval of the activated partial thromboplastin time (aPTT) by the hospital laboratory has generated interest in point-of-care heparin monitoring. The activated clotting time (ACT), the aPTT as assessed by both a new portable device, as well as the hospital laboratory, and heparin levels (H) were obtained from the same sample of blood in 100 patients receiving intravenous heparin. There was an excellent correlation between the aPTT determined at the bedside and by the hospital laboratory (r = .89). The ACT did not correlate well with either the laboratory or bedside aPTT (r = .63, .68 respectively). In the sub-therapeutic and therapeutic range, there was essentially no correlation between ACT and H. Only ACT values > 225 sec were predictive of therapeutic or supra-therapeutic aPTTs. ACT values < 225 sec, however, were not useful in predicting degree of anticoagulation. In situations in which the maintenance of therapeutic anticoagulation is critical as well as those in which the determination of lack of anticoagulation is required, the bedside determination of aPTT appears to be a useful tool.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Reiner et al. (1994) studied this question.

synapsesocial.com/papers/6a70fd48e71d69abee08fa6chttps://doi.org/10.1002/ccd.1810320112