December 1, 1989Pacing and Clinical Electrophysiology11 citations

Sequential Pulse Countershock Between Two Transvenous Catheters: Feasibility, Safety, and Efficacy

View Full Paper
RYRaymond YeeDJDouglas L. JonesGKGeorge J. Klein

Structured PICO

Does sequential pulse countershock delivered through two endocardial catheters effectively terminate VT and VF in patients undergoing electrophysiology studies?

P
Population
34 patients (31 men, 3 women) with a mean age of 56.8 +/- 10.1 years undergoing electrophysiology studies (EPS). Etiologies included ischemic heart disease (n=26), cardiomyopathy (n=4), repaired tetralogy of Fallot (n=1), heart transplant (n=1), and no identifiable heart disease (n=2).
I
Intervention
Sequential pulse countershock (SqCS) delivered solely through two endocardial catheters positioned in the right ventricular apex (RVA) and the coronary sinus (CS).
O
Outcome
Termination of ventricular tachycardia (VT) and ventricular fibrillation (VF)surrogate

Sequential pulse countershock delivered solely between endocardial catheter electrodes is feasible and effective for terminating VT and VF using energy doses within the range of existing implantable cardioverter defibrillators.

Abstract

We evaluated the feasibility, safety, and efficacy of sequential pulse countershock (SqCS) delivered solely through two endocardial catheters for the termination of ventricular tachycardia (VT) and fibrillation (VF) in patients undergoing electrophysiology studies (EPS). Thirty-four patients (31 men, 3 women) with a mean age of 56.8 +/- 10.1 years were studied. Etiology of VT/VF was ischemic heart disease (n = 26), cardiomyopathy (4) repaired tetralogy of Fallot (n = 1), heart transplant (n = 1), and no identifiable heart disease (n = 2). Catheters were positioned successfully in 29 patients. These were positioned in the right ventricular apex (RVA) and the coronary sinus (CS), respectively. The RVA electrode served as the common cathode for both pulses. The two electrodes located near the right atrium/superior vena cava junction served as anode for pulse 1 while the distal CS electrodes served as anode for pulse 2. Twenty-nine induced VT episodes with cycle length (CL) 220-370 msec were treated. SqCS successfully terminated 15 VT (100-500V) while 14 were accelerated or degenerated to VF. VTCL was longer in successful SqCS episodes than in those that were accelerated (285 +/- 17.3 vs 245 +/- 30.8 msec, P less than .003). Of 26 VF episodes, 21 were terminated with SqCS (500-900V) and 5 were terminated by transthoracic rescue shocks. On 2 occasions, failure to defibrillate was attributable to poor catheter position at the time of shock. No complications occurred. We conclude that SqCS delivered solely between endocardial catheter electrodes is feasible and effective using energy doses within the range of existing implantable cardioverter defibrillators.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Yee et al. (1989) studied this question.

synapsesocial.com/papers/6a716d9d3ce530166bc31767https://doi.org/10.1111/j.1540-8159.1989.tb01878.x