Inappropriate Detection of Supraventricular Arrhythmias by Implantable Dual Chamber Defibrillators: A Comparison of Four Different Algorithms

View Full Paper
FHFlorian HintringerInnsbruck Medical UniversitySSSeverin P. SchwarzacherJagiellonian UniversityGEGuenther EiblUniversität Innsbruck

Structured PICO

Do different dual chamber ICD algorithms differ in their specificity for detecting supraventricular tachyarrhythmias?

P
Population
Library of 71 supraventricular and 15 ventricular tachyarrhythmias recorded from patients undergoing invasive electrophysiological studies and catheter ablation
I
Intervention
Four different DDD ICD detection algorithms (Phylax AV [Biotronik], Defender IV [ELA], Ventak AV III DR [Guidant], and Gem DR 7271 [Medtronic])
C
Comparator
Direct comparison among the four algorithms
O
Outcome
Specificity for detection of supraventricular tachyarrhythmiassurrogate

Different dual chamber ICD algorithms vary significantly in their specificity for detecting supraventricular tachyarrhythmias, with ELA and Medtronic algorithms performing better due to more detailed analysis of AV timing.

Abstract

Inappropriate therapy of supraventricular tachyarrhythmias by an ICD is still a common problem. Dual chamber (DDD) ICDs provide additional atrial sensing and should result in higher specificity for detection of supraventricular tachyarrhythmias. However, a direct comparison of different dual chamber algorithms has not been reported. The detection algorithms of four different DDD ICDs were tested: Phylax AV, Defender IV, Ventak AV III DR, and Gem DR 7271. Based on arrhythmias recorded from patients undergoing invasive electrophysiological studies and in many cases of catheter ablation at our institution, a library consisting of 71 supraventricular and 15 ventricular tachyarrhythmias was created. The library consists of episodes of atrial fibrillation, atrial flutter with different AV conduction, typical and atypical AV nodal reentrant tachycardia, AV reentrant tachycardia, sinus tachycardia, and ventricular tachycardia with and without ventriculoatrial conduction. Atrial fibrillation was appropriately classified by all four algorithms. However, the specificity for detection of other supraventricular tachyarrhythmias achieved by the Biotronik (12%) and the Guidant (11%) devices was significantly lower compared to the specificity of the ELA (28%) and the Medtronic DDD ICD (20%). This is due to the fact that the Biotronik and the Guidant algorithm classified all supraventricular tachyarrhythmias resulting in a stable ventricular rate as ventricular tachycardia, whereas the ELA and Medtronic algorithms performed a more detailed analysis by assessment of PR association, atrial onset, or timing of the atrial event relative to the ventricular event, respectively. Atrial fibrillation, the most common supraventricular tachyarrhythmia in patients with ICD, was detected by all devices.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Hintringer et al. (2001) studied this question.

synapsesocial.com/papers/6a728a67febe604dd70aa887https://doi.org/10.1046/j.1460-9592.2001.00835.x
Ask AI
Helpful
Bookmark
Share
View Full Paper