Measurement variation among 12 electronic home blood pressure monitors

View Full Paper
SYSteven A. YarowsUniversity of MichiganRBRobert D. BrookPreventive Cardiology

Structured PICO

Does the use of various electronic home blood pressure monitors result in significant measurement differences compared to the Omron 712c and auscultatory method in normotensive subjects?

P
Population
Normotensive subjects in an ambulatory setting
I
Intervention
Blood pressure measurement using 12 different electronic home blood pressure monitors (Sunbeam 7654, Sunbeam 7623, Omron 711, Omron 432c, A&D-UA767, Lumiscope 1085M, Omron 725CIC, Assure A30, Lumiscope 1083N, Omron 815, Omron 605, and Assure BD-W20)
C
Comparator
Omron 712c electronic home monitor and auscultatory method by trained observers
O
Outcome
Systolic and diastolic blood pressure measurementssurrogate

Small measurement differences among various electronic home blood pressure monitors support their use in both office and home settings.

Limitations

  • Differences in patient composition studied could account for the difference

Abstract

Although there are AAMI and BHS standards for accuracy of electronic home electronic blood pressure monitors (HBPM), patient composition differences and differences in manufacturer's algorithm for calculation of the systolic and diastolic measurement may result in measurement differences between monitors. The aim of this study was the measurement of differences among HBPM. Paired comparisons were performed between the Omron 712c electronic home monitor and each of 12 other HBPM (Sunbeam 7654, Sunbeam 7623, Omron 711, Omron 432c, A&D-UA767, Lumiscope 1085M, Omron 725CIC, Assure A30, Lumiscope 1083N, Omron 815, Omron 605, and Assure BD-W20), in addition to comparison to the auscultatory method by trained observers. Measurements were made in normotensive subjects in an ambulatory setting. The main outcome measures were systolic and diastolic blood pressure measurements. All of the HBPM, except for the Sunbeam 7654 and the Assure A30/ BD-W20 (wrist) models, demonstrated small differences of <4/4 mm Hg for systolic/diastolic measurements with pulse measurement differences of <3 beats/min. These differences were less than the differences previously reported for office BP auscultation of 6/5-10 mm Hg for systolic/diastolic measurements. The Omron 712c, passing previous AAMI and BHS standards, measured the systolic reading within 2 mm Hg of auscultatory mercury or aneroid measurement and under-measured the diastolic by 6-9 mm Hg. Differences in the patient composition studied could account for the difference. The wrist and finger manometers performed clinically similar to the Omron 712c, except for the Assure BD-W20, which overmeasured the diastolic by 7 mm Hg. It is concluded that the small differences among the various HBPM, which are less than those in clinical office auscultation, should encourage greater use of electronic manometers in the office and at home.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Yarows et al. (2000) studied this question.

synapsesocial.com/papers/6a72c4138031ec7bb1dd7bd6https://doi.org/10.1016/s0895-7061(99)00182-x
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: the Omron HEM-722C and HEM-735C1999 · 73 citations
  2. 2Clinical Evaluation of Finger Blood Pressure Measurement Devices for Home-use.1993 · 3 citations
  3. 3Short report: An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices1993 · 524 citations
  4. 4The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems1990 · 776 citations
  5. 5[Distribution and acceptance of self measurement of blood pressure in the Hamburg area].1989 · 7 citations