Reliability of home blood pressure monitoring

View Full Paper
MJMi‐Hyang JungGKGee-Hee KimJKJi-Hoon Kim

Structured PICO

Are automated upper arm-type home blood pressure devices accurate compared to mercury sphygmomanometers in hypertensive patients?

P
Population
212 hypertensive patients using automated upper arm-type blood pressure devices
I
Intervention
Automated upper arm-type home blood pressure devices
C
Comparator
Mercury sphygmomanometer
O
Outcome
Accuracy of the device (defined as less than a 5 mmHg difference in mean BP values compared to mercury sphygmomanometer)surrogate

Nonvalidated home blood pressure devices are widely used and significantly more likely to be inaccurate than validated devices, emphasizing the need for clinical verification of device accuracy.

Abstract

OBJECTIVE: Home blood pressure (BP) monitoring offers clinically relevant information enriched with more abundant data. However, there are few studies addressing the reliability of home BP devices and the quality of its data. This study aimed to evaluate the current status of home BP devices in terms of validation and accuracy. PARTICIPANTS AND METHODS: Hypertensive patients with automated upper arm-type BP devices were consecutively enrolled. First, the validation status of each device was assessed through the website. Next, the accuracy of the individual device was evaluated by comparing the mean BP values between the automated device and a mercury sphygmomanometer. Accuracy of the device was defined as having less than a 5 mmHg difference in the mean BP values. RESULTS: A total of 212 individuals were analyzed; 38.7% (82 of 212) of the participants used validated devices and 85.4% (181 of 212) were accurate. Inaccuracy was more common with nonvalidated devices than validated devices 19.2% (n=25) vs. 7.3% (n=6), P=0.017. The range of inaccuracy of the validated devices was 6-26 mmHg for the systolic BP and 6-11 mmHg for the diastolic BP. CONCLUSION: The present study showed that nonvalidated devices are used widely in clinical practice and a substantial portion is inaccurate. Therefore, recommendation of validated devices should be the first step. Furthermore, all devices need to be examined for accuracy before use irrespective of their validation status.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Jung et al. (2015) studied this question.

synapsesocial.com/papers/6a7dcd02f9fad9e96bb873f5https://doi.org/10.1097/mbp.0000000000000121
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Inaccuracy of home sphygmomanometers2011 · 27 citations
  2. 2Factors Influencing the Accuracy of Home Blood Pressure Measurement1997 · 16 citations
  3. 3Self‐measurement of blood pressure at home: Is it reliable?2008 · 17 citations
  4. 4Comparison of two validated oscillometric devices in a home-like setup reveals pronounced blood pressure differences and reduced precision2026
  5. 5Home sphygmomanometers: validation versus accuracy2009 · 48 citations