Assessment of structural valve deterioration of transcatheter aortic bioprosthetic balloon-expandable valves using the new European consensus definition

View Full Paper
HEHélène EltchaninoffÉDÉric DurandGAG. Avinée

Key Result

Transcatheter aortic bioprosthetic valves demonstrated an 8-year incidence of structural valve deterioration of 3.2% (95% CI: 1.45-6.11) and bioprosthetic valve failure of 0.58%.

Study Design

Type

Cohort (n=378)

Structured PICO

P
Population
378 TAVI patients implanted up to September 2012 with a minimal theoretical 5-year follow-up, with a mean age of 83.3 years.
E
Exposure
Transcatheter aortic valve implantation (TAVI) with balloon-expandable bioprosthetic valves
O
Outcome
Structural transcatheter valve deterioration (SVD) and bioprosthetic valve failure (BVF) using new European consensus definitionshard clinical

Long-term follow-up of early TAVI patients demonstrates low rates of structural valve deterioration and bioprosthetic valve failure at 8 years, though limited by high competing mortality.

Limitations

  • poor survival of the very high-risk/compassionate early population

Abstract

AIMS: Durability of transcatheter aortic bioprosthetic valves remains a major issue. Standardised definitions of deterioration and failure of bioprosthetic valves have recently been proposed. The aim of this study was to assess structural transcatheter valve deterioration (SVD) and bioprosthetic valve failure (BVF) using these new definitions. METHODS AND RESULTS: All TAVI patients implanted up to September 2012 with a minimal theoretical five-year follow-up were included. Systematic clinical and echocardiographic follow-up was performed annually. New standardised definitions were used to assess durability of transcatheter aortic bioprosthetic valves. From 2002 to 2012, 378 patients were included. Mean age and logistic EuroSCORE were 83.3±6.8 years and 22.8±13.1%. Thirty-day mortality was 13.2%. Nine patients had SVD including two severe forms and two patients had definite late BVF. The incidence of SVD and BVF at eight years was 3.2% (95% CI: 1.45-6.11) and 0.58% (95% CI: 0.15-2.75), respectively. CONCLUSIONS: Even though limited by the poor survival of the very high-risk/compassionate early population, our data do not demonstrate any alarm concerning transcatheter aortic valve durability. Careful prospective assessment in younger and lower-risk patients and comparison with surgical bioprosthetic valves are required for further assessment of the long-term durability of transcatheter valves.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Eltchaninoff et al. (2018) conducted a cohort in Transcatheter aortic bioprosthetic valves (n=378). Transcatheter aortic bioprosthetic balloon-expandable valves was evaluated on Structural transcatheter valve deterioration (SVD) at eight years (95% CI 1.45-6.11). Transcatheter aortic bioprosthetic valves demonstrated an 8-year incidence of structural valve deterioration of 3.2% (95% CI: 1.45-6.11) and bioprosthetic valve failure of 0.58%.