A Fast, Reliable, Automatic Shimming Procedure Using 1H Chemical-Shift-Imaging Spectroscopy

View Full Paper
JHJiani HuTJTariq JavaidFAFernando Arias‐Mendoza

Key Points

Key points are not available for this paper at this time.

Abstract

The importance and the difficulty of achieving good B0 homogeneity over the volume of interest in in vivo NMR spectroscopy are well known. A fast, reliable, versatile, and fully automatic shimming procedure has been developed, using 3D chemical-shift imaging to measure the field distribution using only the water peak in the sample of interest. The procedure minimizes the mean-square error in the field distribution with respect to a constant field where the signal exists (e.g., the head). This produces a set of linear equations that can be solved by standard matrix methods. The procedure has been applied on a commercial imager producing water linewidths from the entire head as low as 8 Hz at 1.5 T. The mean linewidth and standard deviation from 94 head studies were 11.7 +/- 1.9 Hz.