October 14, 2010Journal of Cardiovascular Translational Research84 citationsOpen Access

Direct Left Atrial Pressure Monitoring in Severe Heart Failure: Long-Term Sensor Performance

RTRichard W. TroughtonJRJay RitzemaNENeal Eigler

Structured PICO

Does an implantable left atrial pressure monitoring system provide accurate and stable long-term hemodynamic measurements in patients with advanced heart failure?

P
Population
84 patients with advanced/chronic severe heart failure
I
Intervention
Implantable left atrial pressure (LAP) monitoring system (HeartPOD) via percutaneous transseptal implantation
O
Outcome
Sensor stability (drift), accuracy (correlation with simultaneous pulmonary capillary wedge pressure), and freedom from device failure/anomalies at up to 48 monthssurrogate

Direct left atrial pressure monitoring using a permanently implanted sensor is feasible, accurate, and stable over the long term in patients with advanced heart failure.

Abstract

We report the stability, accuracy, and development history of a new left atrial pressure (LAP) sensing system in ambulatory heart failure (HF) patients. A total of 84 patients with advanced HF underwent percutaneous transseptal implantation of the pressure sensor. Quarterly noninvasive calibration by modified Valsalva maneuver was achieved in all patients, and 96.5% of calibration sessions were successful with a reproducibility of 1.2 mmHg. Absolute sensor drift was maximal after 3 months at 4.7 mmHg (95% CI, 3.2-6.2 mmHg) and remained stable through 48 months. LAP was highly correlated with simultaneous pulmonary wedge pressure at 3 and 12 months (r = 0.98, average difference of 0.8 ± 4.0 mmHg). Freedom from device failure was 95% (n = 37) at 2 years and 88% (n = 12) at 4 years. Causes of failure were identified and mitigated with 100% freedom from device failure and less severe anomalies in the last 41 consecutive patients (p = 0.005). Accurate and reliable LAP measurement using a chronic implanted monitoring system is safe and feasible in patients with advanced heart failure.