December 1, 1956Review of Scientific Instruments13 citations

Rotating Electrometer for Comparative Work Function Measurements

View Full Paper
HKHenry H. Kolm

Key Points

Key points are not available for this paper at this time.

Abstract

An apparatus for measuring work functions by the Kelvin method is described in which the customary vibrating motion is replaced by rotary motion. A set of three coplanar samples is mounted on a turntable which rotates below a set of three stationary samples. This arrangement permits simultaneous observation of three electrode pairs in all permutations and provides ready access to the sample surfaces for purposes of treatment, illumination, inspection, etc., without the necessity for repositioning. The rigid structure also permits the application of electrostatic fields between electrodes, which is difficult in a vibrating system. Measurements can be made in high vacuum by driving the turntable inductively with coils located outside the glass envelope. The instrument has a sensitivity of ±1 millivolt.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Henry H. Kolm (1956) studied this question.

synapsesocial.com/papers/6a827c08fe5ce613f33ad238https://doi.org/10.1063/1.1715450