April 29, 2023Archives of cardiovascular diseases22 citationsOpen Access

Clinical validation of an artificial intelligence algorithm offering cross-platform detection of atrial fibrillation using smart device electrocardiograms

View Full Paper
DMDiego MannhartBLBaptiste LefebvreCGChristophe Gardella

Structured PICO

Does a novel deep neural network-based algorithm reduce inconclusive tracings and maintain diagnostic accuracy for atrial fibrillation detection compared to manufacturers' algorithms on smart devices?

P
Population
117 patients, median age 65 years, 27% female, with atrial fibrillation present at time of recording in 30%.
I
Intervention
Novel deep neural network-based algorithm for detecting atrial fibrillation from single-lead ECGs on five smart devices (Apple, Fitbit, AliveCor, Samsung, Withings).
C
Comparator
Manufacturers' algorithms of the five smart devices, using physician-interpreted 12-lead ECG as the reference standard.
O
Outcome
Rate of inconclusive tracings and diagnostic accuracy for the detection of atrial fibrillation.surrogate

A novel deep neural network-based algorithm significantly reduces inconclusive ECG tracings from smart devices while maintaining high diagnostic accuracy for atrial fibrillation detection.

Abstract

BACKGROUND: Several smart devices are able to detect atrial fibrillation automatically by recording a single-lead electrocardiogram, and have created a work overload at the hospital level as a result of the need for over-reads by physicians. AIM: To compare the atrial fibrillation detection performances of the manufacturers' algorithms of five smart devices and a novel deep neural network-based algorithm. METHODS: We compared the rate of inconclusive tracings and the diagnostic accuracy for the detection of atrial fibrillation between the manufacturers' algorithms and the deep neural network-based algorithm on five smart devices, using a physician-interpreted 12-lead electrocardiogram as the reference standard. RESULTS: Of the 117 patients (27% female, median age 65 years, atrial fibrillation present at time of recording in 30%) included in the final analysis (resulting in 585 analyzed single-lead electrocardiogram tracings), the deep neural network-based algorithm exhibited a higher conclusive rate relative to the manufacturer algorithm for all five models: 98% vs. 84% for Apple; 99% vs. 81% for Fitbit; 96% vs. 77% for AliveCor; 99% vs. 85% for Samsung; and 97% vs. 74% for Withings (P<0.01, for each model). When applying our deep neural network-based algorithm, sensitivity and specificity to correctly identify atrial fibrillation were not significantly different for all assessed smart devices. CONCLUSION: In this clinical validation, the deep neural network-based algorithm significantly reduced the number of tracings labeled inconclusive, while demonstrating similarly high diagnostic accuracy for the detection of atrial fibrillation, thereby providing a possible solution to the data surge created by these smart devices.