April 1, 1975Metrologia9 citations

The Viscosity Scale of the NPL

View Full Paper
JDJ E Daborn

Key Points

Key points are not available for this paper at this time.

Abstract

The establishment of the NPL viscosity scale using glass capillary viscometers is described and the values obtained for NPL viscometers on subsequent recalibrations over a period of some years are given. The results of comparisons with independent viscosity scales in other countries are also tabulated. From the results obtained on recalibration of the NPL viscometers and the comparisons made with other independent laboratories, it is concluded that the present scale is realized with an uncertainty of about 0.2% at low viscosities (5 cSt) rising to about 0.5% at high viscosities (5000 cSt).

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

J E Daborn (1975) studied this question.