January 1, 2005Acta Orthopaedica391 citationsOpen Access

Pelvic tilt makes acetabular cup navigation inaccurate

View Full Paper
BLBurkhard LembeckOMOtto MuellerPRPatrik Reize

Key Points

Key points are not available for this paper at this time.

Abstract

BACKGROUND: Modern navigation techniques allow precise positioning of the acetabular cup relative to the anterior pelvic plane. Variations in pelvic tilt will affect the resulting spatial orientation of the cup. METHODS: We measured pelvic tilt in 30 volunteers with an inclinometer combined with an ultrasonographic position measurement system. A mathematical algorithm was developed to calculate the resulting cup position measured on standard radiographs, depending on pelvic tilt. RESULTS: Average pelvic tilt at rest was -4 degrees in the lying position and -8 degrees in the standing position, and ranged from -27 degrees to +3 degrees. Pelvic reclination of 1 degree will lead to functional anteversion of the cup of approximately 0.7 degree. INTERPRETATION: Pelvic tilt makes navigation systems referring to the anterior plane inaccurate.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Lembeck et al. (2005) studied this question.

synapsesocial.com/papers/6a8431d8dc9bcdcbf8c6bbf7https://doi.org/10.1080/17453670510041501