PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
September 30, 2020Magnetic Resonance in Medicine22 citationsOpen Access

Fully self‐gated whole‐heart 4D flow imaging from a 5‐minute scan

Structured PICO

Does ReVEAL4D reconstruction of a 5-minute self-gated 4D flow MRI scan provide accurate flow quantification compared to 2D phase-contrast MRI?

P
Population
8 healthy volunteers and 2 patients (n=10)
I
Intervention
ReVEAL4D reconstruction of highly accelerated 5-minute Cartesian acquisition with self-gating
C
Comparator
2D phase-contrast MRI (2D-PC), L1-SENSE, and traditional parallel MRI-based 4D flow imaging
O
Outcome
Flow quantification accuracy (net flow, peak flow rate, peak velocity)surrogate

ReVEAL4D enables accurate whole-heart 4D flow imaging from a 5-minute scan, overcoming underestimation issues seen with L1-SENSE.

Abstract

PURPOSE: To develop and validate an acquisition and processing technique that enables fully self-gated 4D flow imaging with whole-heart coverage in a fixed 5-minute scan. THEORY AND METHODS: The data are acquired continuously using Cartesian sampling and sorted into respiratory and cardiac bins using the self-gating signal. The reconstruction is performed using a recently proposed Bayesian method called ReVEAL4D. ReVEAL4D is validated using data from 8 healthy volunteers and 2 patients and compared with compressed sensing technique, L1-SENSE. RESULTS: Healthy subjects-Compared with 2D phase-contrast MRI (2D-PC), flow quantification from ReVEAL4D shows no significant bias. In contrast, the peak velocity and peak flow rate for L1-SENSE are significantly underestimated. Compared with traditional parallel MRI-based 4D flow imaging, ReVEAL4D demonstrates small but significant biases in net flow and peak flow rate, with no significant bias in peak velocity. All 3 indices are significantly and more markedly underestimated by L1-SENSE. Patients-Flow quantification from ReVEAL4D agrees well with the 2D-PC reference. In contrast, L1-SENSE markedly underestimated peak velocity. CONCLUSIONS: The combination of highly accelerated 5-minute Cartesian acquisition, self-gating, and ReVEAL4D enables whole-heart 4D flow imaging with accurate flow quantification.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

A 2020 study studied this question.

synapsesocial.com/papers/6a87d0b502f504243ce964c4https://doi.org/10.1002/mrm.28491
Ask AI
Helpful
Bookmark
Share
View Full Paper