A glass disk calorimeter for pulsed lasers

JEJ. G. Edwards

Key Points

Key points are not available for this paper at this time.

Abstract

Because existing calorimeters are not adequate to measure Q -switched laser pulses accurately, we have designed a calorimeter in which the beam passes through two glass disks, one of green plate (absorbing from 0·6-1·6 μm), the other of clear plate. A laser pulse creates a temperature difference which is sensed by thermocouples. 70% of the incident energy passes through the device and is available for experiment. Thermal drift is low enough to obtain 1% accuracy at 10 mJ input energy. For energies up to 0·1 J, any pulse length from 1 s to 0·1 ps can be used. Energies up to 50 J can be used provided the power density at the glass does not exceed 40 MW cm −2 . The device is dependent on a conventional calorimeter for calibration (limiting the absolute accuracy to 2%), and the sensitivity is wavelength dependent, thus limiting its use to wavelengths where a free running laser is available; ways of overcoming this disadvantage are discussed.