December 1, 1975Clinical Chemistry78 citations

Gas-Chromatographic Analysis for Cocaine in Human Plasma, with Use of a Nitrogen Detector

View Full Paper
PJPeter JatlowDBDavid N. Bailey

Key Points

Key points are not available for this paper at this time.

Abstract

We describe a procedure for measuring the concentration of cocaine in plasma by gas chromatography, with use of a nitrogen detector. The procedure was tested on patients receiving intranasal cocaine for topical anesthesia. The propyl ester of benzoylecgonine is used as an internal standard. The lower limit of sensitivity of the procedure is 5--10 mug of cocaine per liter of plasma. The coefficients of variation are 5.3% and 3.0% for concentrations of 50 and 100 mug/liter, respectively.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Jatlow et al. (1975) studied this question.

synapsesocial.com/papers/6a904bc57b6e700e2c8a4b87https://doi.org/10.1093/clinchem/21.13.1918