June 1, 1970Journal of Chromatographic Science43 citationsOpen Access

Curie Point Pyrolysis Gas Chromatography

View Full Paper
CBC. R. BuhlerWSW. Simon

Key Points

Key points are not available for this paper at this time.

Abstract

In Curie Point pyrolysis gas chromatography a ferromagnetic conductor in contact with the sample is heated inside a low volume glass capillary which is inserted into an rf coil with the carrier gas flowing through the capillary. The intrinsic properties of the conductor permit a self-control of the temperature to the Curie Point. The optimum wire diameter for a rapid warm-up is a function of the rf frequency. In order to have a fast temperature drop after cutting the rf field, small wire diameters are preferable. For such conductors relatively high frequencies are inherently necessary for a fast stabilization to the Curie Point. So far no differences in the self-controlled end-temperatures have been detected for wires of different origin.