First-in-human evaluation of a novel balloon-expandable transcatheter heart valve in patients with severe symptomatic native aortic stenosis: the MyVal-1 study

View Full Paper
SSSamin K. SharmaRRRavinder Singh RaoPCPraveen Chandra

Key Result

Implantation of the Myval transcatheter heart valve significantly improved mean aortic valve gradient from 47.4 to 12.0 mmHg (p<0.0001) at 12 months, with no new permanent pacemaker requirement.

Structured PICO

Does the Myval transcatheter heart valve improve clinical and echocardiographic outcomes safely in intermediate- or high-risk patients with severe symptomatic native aortic stenosis?

P
Population
30 intermediate- or high-risk patients with severe symptomatic native aortic stenosis, followed for 12 months.
I
Intervention
Implantation of the next-generation balloon-expandable Myval transcatheter heart valve (THV).
O
Outcome
Safety and efficacy (including NYHA class, 6-minute walk test, KCCQ score, echocardiographic parameters, mortality, and pacemaker requirement) at 30-day and 12-month follow-up.safety

The first-in-human MyVal-1 study demonstrated that the Myval transcatheter heart valve is safe and effective at 12 months for severe symptomatic aortic stenosis, notably with a 0% new permanent pacemaker rate.

Limitations

  • Small number of patients
  • Lack of long-term follow-up
  • small sample size
  • short follow-up

Abstract

AIMS: The aim of this study was to demonstrate the safety and efficacy of the next-generation balloon-expandable Myval transcatheter heart valve (THV) in an intermediate- or high-risk patient population with severe symptomatic native aortic stenosis. METHODS AND RESULTS: MyVal-1 was a first-in-human, prospective, multicentre, single-arm, open-label study. Between June 2017 and February 2018, a total of 30 patients were enrolled at 14 sites across India. Mean age was 75.5±6.7 years; 43.3% had coronary artery disease. The mean Society of Thoracic Surgeons score was 6.4±1.8% and 100% of the patients were in New York Heart Association (NYHA) functional Class II/III/IV pre-procedure. The six-minute walk test and Kansas City Cardiomyopathy Questionnaire (KCCQ) scores were recorded. After successful implantation of the Myval THV, 96.6% and 100% were in NYHA functional Class I/II at 30-day and 12-month follow-up, respectively. Outcomes of the six-minute walk test (148.0±87.4 vs 336.0±202.9 m) and KCCQ score (36.6±11.0 vs 65.9±11.4) improved from baseline to 12-month follow-up. The effective orifice area (0.6±0.2 vs 1.8±0.3 cm2, p<0.0001), mean aortic valve gradient (47.4±8.8 vs 12.0±3.3 mmHg, p<0.0001), peak aortic valve gradient (71.7±13.0 vs 20.3±5.9 mmHg, p<0.0001) and transaortic velocity (4.5±0.4 vs 2.2±0.4 m/s, p<0.0001) improved substantially from baseline to 12 months post procedure. Four all-cause mortality cases were reported up to 12 months. Moreover, there was no other moderate/severe paravalvular leak, aortic regurgitation or need for new permanent pacemaker (PPM) up to 12-month follow-up. CONCLUSIONS: The MyVal-1 study demonstrated the primary safety and efficacy of the Myval THV with no new PPM requirement up to 12-month follow-up. However, future trials with a larger number of patients and long-term follow-up are warranted to establish the safety and efficacy of the device.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Sharma et al. (2020) studied severe symptomatic native aortic stenosis (n=30). Myval transcatheter heart valve was evaluated on Safety and efficacy. Implantation of the Myval transcatheter heart valve significantly improved mean aortic valve gradient from 47.4 to 12.0 mmHg (p<0.0001) at 12 months, with no new permanent pacemaker requirement.