Comparison of the Hemodynamic Performance of Percutaneous and Surgical Bioprostheses for the Treatment of Severe Aortic Stenosis

View Full Paper
MCMarie‐Annick ClavelJWJohn G. WebbPPPhilippe Pîbarot

Key Result

Percutaneous aortic valve implantation yielded a lower mean transprosthetic gradient at discharge (10 vs 13 and 14 mm Hg; P<0.001) than surgical valves, but higher rates of aortic regurgitation.

Key Points

  • To evaluate and compare the hemodynamic performance of catheter-implanted bioprosthetic aortic valves against stented and stentless surgical bioprostheses in patients with severe aortic stenosis.
  • Matched cohort comparison of 50 patients undergoing percutaneous aortic valve implantation (PAVI) matched 1:1 for sex, aortic annulus diameter, left ventricular ejection fraction, body surface area, and body mass index with 50 stented (SAVR-ST) and 50 stentless (SAVR-SL) surgical aortic valve replacement patients (total N=150).
  • Doppler echocardiographic evaluations were prospectively conducted before intervention, at hospital discharge, and at 6- to 12-month follow-up.
  • Mean transprosthetic gradient at discharge was significantly lower in the PAVI group (10 ± 4 mm Hg) than in the SAVR-ST (13 ± 5 mm Hg) and SAVR-SL (14 ± 6 mm Hg) groups (p < 0.001), remaining lower than SAVR-ST at follow-up (p < 0.001).
  • Severe prosthesis-patient mismatch occurred in 6% of PAVI patients versus 28% in the SAVR-ST and 20% in the SAVR-SL cohorts (p = 0.007).
  • Aortic regurgitation occurred more frequently with PAVI (mild: 42%, moderate: 8%) than with SAVR-ST (mild: 10%, moderate: 0%) and SAVR-SL (mild: 12%, moderate: 0%) at discharge (p < 0.0001) and remained higher at follow-up (p < 0.0001).

Study Design

Type

Cohort (n=150)

PICO

P
Population
150 patients with severe aortic stenosis undergoing percutaneous or surgical aortic valve replacement, matched 1:1 for baseline characteristics, followed for 6 to 12 months.
E
Exposure / Comparator
Percutaneous aortic valve implantation (PAVI) vs Surgical aortic valve replacement (SAVR) with stented or stentless bioprostheses
O
Primary Outcome
Mean transprosthetic gradient at discharge, p=< 0.001

Main Result

Absolute Event Rate: 10% vs 13%

p-value: p=< 0.001

Abstract

OBJECTIVES: This study was undertaken to compare the hemodynamic performance of a percutaneous bioprosthesis to that of surgically implanted (stented and stentless) bioprostheses for the treatment of severe aortic stenosis. METHODS: Fifty patients who underwent percutaneous aortic valve implantation (PAVI) with the Cribier-Edwards or Edwards SAPIEN bioprosthetic valve (Edwards Lifesciences, Inc., Irvine, California) were matched 1:1 for sex, aortic annulus diameter, left ventricular ejection fraction, body surface area, and body mass index, with 2 groups of 50 patients who underwent surgical aortic valve replacement (SAVR) with a stented valve (Edwards Perimount Magna SAVR-ST group), or a stentless valve (Medtronic Freestyle, Medtronic, Minneapolis, Minnesota SAVR-SL group). Doppler echocardiographic data were prospectively obtained before the intervention, at discharge, and at 6- to 12-month follow-up. RESULTS: Mean transprosthetic gradient at discharge was lower (p < 0.001) in the PAVI group (10 +/- 4 mm Hg) compared with the SAVR-ST (13 +/- 5 mm Hg) and SAVR-SL (14 +/- 6 mm Hg) groups. Aortic regurgitation (AR) occurred more frequently in the PAVI group (mild: 42%, moderate: 8%) compared with the SAVR-ST (mild: 10%, moderate: 0%) and SAVR-SL (mild: 12%, moderate: 0%) groups (p < 0.0001). At follow-up, the mean gradient in the PAVI group remained lower (p < 0.001) than that of the SAVR-ST group, but was similar to that of the SAVR-SL group. The incidence of severe prosthesis-patient mismatch was significantly lower (p = 0.007) in the PAVI group (6%) compared with the SAVR-ST (28%) and SAVR-SL (20%) groups. However, the incidence of AR remained higher (p < 0.0001) in the PAVI group compared with the 2 other groups. CONCLUSIONS: PAVI provided superior hemodynamic performance compared with the surgical bioprostheses in terms of transprosthetic gradient and prevention of severe prosthesis-patient mismatch, but was associated with a higher incidence of AR.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Clavel et al. (2009) conducted a cohort in Severe aortic stenosis (n=150). Percutaneous aortic valve implantation (PAVI) vs. Surgical aortic valve replacement (SAVR) with stented or stentless bioprostheses was evaluated on Mean transprosthetic gradient at discharge (p=< 0.001). Percutaneous aortic valve implantation yielded a lower mean transprosthetic gradient at discharge (10 vs 13 and 14 mm Hg; P<0.001) than surgical valves, but higher rates of aortic regurgitation.

synapsesocial.com/papers/6aa41bd983096b243c9c9ca2https://doi.org/10.1016/j.jacc.2009.01.060
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Genetic Differences in Endothelial Cells May Determine Atherosclerosis Susceptibility2000 · 60 citations
  2. 2Effects of Off-Pump Versus On-Pump Coronary Surgery on Reversible and Irreversible Myocardial Injury2004 · 187 citations
  3. 3Does the Use of Stentless Aortic Valves in a Subcoronary Position Prevent Patient-Prosthesis Mismatch for Small Aortic Annulus?2008 · 19 citations
  4. 4Prosthesis-patient mismatch: definition, clinical impact, and prevention2005 · 607 citations
  5. 5Impact of Prosthesis-Patient Mismatch on Cardiac Events and Midterm Mortality After Aortic Valve Replacement in Patients With Pure Aortic Stenosis2006 · 223 citations