The wrist monitor overdiagnosed hypertension, identifying nearly twice as many hypertensive individuals compared to the upper-arm device.
Observational (n=121)
No
Does a wrist-worn oscillometric blood pressure monitor provide accurate blood pressure measurements and hypertension classification compared to an upper-arm monitor when placed by patients in a home-like setup?
Wrist-worn blood pressure monitors exhibit significant bias, reduced precision, and tend to overdiagnose hypertension compared to upper-arm monitors when used by patients in home-like settings.
Effect estimate: null (95% CI null)
Absolute Event Rate: 47.1% vs 28.1%
p-value: p=<0.0001
Abstract Home blood pressure monitoring (HBPM) is essential for long-term hypertension management, but its accuracy and reliability is questionable due to user inconsistencies and non-standard usage conditions. This study compared two validated HBPM devices – a wrist-based and a reference upper-arm monitor – in a home-like setup where participants placed the devices themselves. A total of 121 participants underwent four concurrent blood pressure (BP) measurements, with two taken in each arm, followed by a crossover of devices between arms after the initial two readings. Wrist-derived BP readings were higher than upper-arm measurements. With higher blood pressure levels, both devices exhibited greater bias, accompanied by reduced precision in systolic BP measurements. Additionally, hypertension classification showed only moderate agreement (Cohen’s kappa=0.58). The wrist monitor tended to over-diagnose hypertension and exhibited greater variability than the upper-arm device. These findings highlight the need for more guidance and education as well as critical evaluation of home blood pressure measurements provided by patients.
Zahnd et al. (2026) conducted an observational in Hypertension (n=121). Omron BP RS4 (wrist monitor) vs. Microlife WatchBP Home (upper-arm monitor) was evaluated on Systolic blood pressure difference between wrist and upper-arm monitors (null, 95% CI null, p=<0.0001). The wrist monitor overdiagnosed hypertension, identifying nearly twice as many hypertensive individuals compared to the upper-arm device.