Magnetic resonance imaging using a pre-specified protocol in 142 CIED patients, including 46 with traditional contraindications, resulted in 0 major complications.
Observational (n=142)
Is magnetic resonance imaging safe in patients with cardiac implantable electronic devices who have traditional contraindications?
MRI can be performed safely using a pre-specified protocol in CIED patients with traditional contraindications such as abandoned leads or depleted batteries, though careful monitoring is required for devices on advisory.
AIMS: Magnetic resonance imaging (MRI) has been reported to be safe in patients with cardiac implantable electronic devices (CIED) provided a specific protocol is followed. The objective of this study was to assess whether this is also true for patients excluded from published protocols. METHODS AND RESULTS: A total of 160 MRIs were obtained in 142 consecutive patients with CIEDs 106 patients had an implantable cardioverter defibrillator (ICD) and 36 had a pacemaker implanted using an adapted, pre-specified protocol. A cardiac MRI was performed in 95 patients, and a spinal/brain MRI was performed in 47 patients. Forty-six patients (32%) had either abandoned leads (n = 10), and/or were pacemaker dependent with an implanted ICD (n = 19), had recently implanted CIEDs (n = 1), and/or had a CIED device with battery depletion (n = 2), and/or a component of the CIED was recalled or on advisory (n = 32). No major complications occurred. Some device parameters changed slightly, but significantly, right after or at 1-week post-MRI without requiring any reprogramming. In one patient with an ICD on advisory, the pacing rate changed inexplicably during one of his two MRIs from 90 to 50 b.p.m. CONCLUSION: Using a pre-specified protocol, cardiac and non-cardiac MRIs were performed in CIED patients with pacemaker dependency, abandoned leads, or depleted batteries without occurrence of major adverse events. Patients with devices on advisory need to be monitored carefully during MRI, especially if they are pacemaker dependent.
Horwood et al. (Thu,) conducted a observational in Cardiac implantable electronic devices (CIEDs) (n=142). Magnetic resonance imaging (MRI) was evaluated on Major complications. Magnetic resonance imaging using a pre-specified protocol in 142 CIED patients, including 46 with traditional contraindications, resulted in 0 major complications.