3D delayed enhancement magnetic resonance sequences improve conducting channel delineation prior to ventricular tachycardia ablation

View Full Paper
DADavid AndreuJOJose T. Ortiz‐PérezJFJuan Fernández‐Armenta

Key Result

CMR-3D imaging improved the delineation of conducting channels prior to VT ablation compared to CMR-2D, identifying a mean of 2.06 vs 1.06 channels, which more closely matched electroanatomic mapping.

Study Design

Type

Observational (n=30)

Structured PICO

Does CMR-3D improve the delineation of conducting channels compared to CMR-2D in patients with structural heart disease referred for VT ablation?

P
Population
30 consecutive patients with structural heart disease referred for ventricular tachycardia (VT) ablation
I
Intervention
Free breathing 3D delayed enhanced navigator gated sequence (CMR-3D) in a 3T-scanner (spatial resolution 1.4 x 1.4 x 1.4 mm)
C
Comparator
Delayed enhanced cardiac magnetic resonance using 2D sequences (CMR-2D) (spatial resolution 1.4 x 1.4 x 5 mm)
O
Outcome
Number of conducting channels (CC) identified compared to those depicted in the electroanatomic map (EAM)surrogate

CMR-3D imaging provides higher spatial resolution and improves the delineation of conducting channels prior to VT ablation compared to standard CMR-2D, despite having lower signal-to-noise and contrast-to-noise ratios.

Main Result

Absolute Event Rate: 2.06% vs 1.06%

Abstract

Introduction: Delayed enhanced cardiac magnetic resonance using 2D sequences (CMR-2D) represents the standard for clinical characterization of myocardium. However, 2D images suffer from insufficient spatial resolution in the Z-axis and missregistration that preclude appropriate visualization of conducting channels (CC), which are critical for ventricular tachycardia (VT) ablation. We studied whether characterization using a free breathing 3D delayed enhanced navigator gated sequence (CMR-3D) proved to be superior than CMR-2D sequences. Methods: We included 30 consecutive patients with structural heart disease referred for VT ablation. Myocardial characterization was conducted prior to the ablation procedure in a 3T-scanner using CMR-2D and CMR-3D sequences, yielding a spatial resolution of 1.4 x 1.4 x 5 mm and 1.4 x 1.4 x 1.4 mm respectively. The scar components, core and border zone (BZ), were quantified in both datasets using the 60% and 40% threshold of maximum pixel intensity, respectively. The signal to noise ratio (SNR) and contrast to noise ratio (CNR) was analyzed in a representative slice in both sequences. For both set of images a 3D reconstruction of the scar was obtained and an electrophysiologyst identified potential CC and compared them to those depicted in the electroanatomic map (EAM). Results: We found no significant differences between the mass of the scar core in the 2D and 3D sequence (mean 7.33±6.56 vs 6.74±5.22 grams respectivelly, p=0.612). However, the BZ mass was smaller in the 2D than in the 3D sequence (9.38±7.72 vs 12.10±7.81 grams; p=0.019). The SNR was higher in the 2D dataset as compared to the 3D (42.93±27.34 vs 28.09±12.85; p=0.005) as well as the CNR (37.01±26.20 vs 20.15±11.41; p=0.001). The mean number of CC identified in the 3D dataset was higher than in the 2D, and closer to those depicted in the EAM (1.06±0.75 for CMR-2D vs 2.63±1.41 for EAM, p=0.005; and 2.06±1.25 for CMR-3D vs 2.63±1.41 for EAM p=0.088). Conclusion: Despite the higher SNR and CNR of standard CMR-2D images, the higher spatial resolution provided by CMR-3D images improved the overall delineation of CC prior to VT ablation.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Andreu et al. (2013) conducted an observational in Structural heart disease referred for VT ablation (n=30). CMR-3D (free breathing 3D delayed enhanced navigator gated sequence) vs. CMR-2D (delayed enhanced cardiac magnetic resonance using 2D sequences) was evaluated on Mean number of conducting channels (CC) identified. CMR-3D imaging improved the delineation of conducting channels prior to VT ablation compared to CMR-2D, identifying a mean of 2.06 vs 1.06 channels, which more closely matched electroanatomic mapping.