November 1, 1997Critical Care Medicine193 citations

Lithium dilution cardiac output measurement

View Full Paper
RLR. A. F. LintonMinistry of DefenceDBDavid BandUniversity of ZambiaMJMax JonasSt Thomas' Hospital

Structured PICO

Does lithium dilution provide accurate cardiac output measurements compared to thermodilution in intensive care and postoperative patients?

P
Population
40 patients in a high-dependency postoperative unit and intensive care unit, including 34 who had undergone heart surgery requiring cardiopulmonary bypass within the previous 2 days, 2 with myocardial infarct, 2 with septicemia, and others with adult respiratory distress syndrome and pericardectomy.
I
Intervention
Lithium dilution cardiac output measurement (single measurement)
C
Comparator
Bolus thermodilution (series of three to six measurements, taking the average of the closest three); continuous thermodilution was also used in a subgroup of 14 patients
O
Outcome
Agreement and difference between cardiac output measurements obtained by lithium dilution and thermodilutionsurrogate

Lithium dilution is an accurate alternative to bolus thermodilution for measuring cardiac output, offering the advantage of not requiring pulmonary artery catheterization.

Abstract

OBJECTIVE: To compare the results of cardiac output measurements obtained by lithium dilution and thermodilution. DESIGN: Case series, observational study. SETTING: High-dependency postoperative unit and intensive care unit of a teaching hospital. PATIENTS: Forty patients were studied. Thirty-four patients had undergoing heart surgery requiring cardiopulmonary bypass within the previous 2 days; the diagnoses in the other patients were myocardial infarct (n = 2), septicemia (n = 2), adult respiratory distress syndrome, and pericardectomy. INTERVENTIONS: Cardiac output was measured five times in each patient, using lithium dilution (single measurement) and bolus thermodilution (series of three to six measurements according to standard clinical practice, taking the average of the closest three). In a subgroup of 14 patients, cardiac output was also measured using "continuous thermodilution." MEASUREMENTS AND MAIN RESULTS: Comparing lithium dilution with bolus thermodilution, the mean of the differences (lithium dilution-thermodilution) was -0.25 +/- 0.46 SD L/min. Linear regression analysis gave y = 0.31 + 0.89x (r2 = .94) for lithium dilution vs. thermodilution. CONCLUSIONS: The overall agreement between the two methods was good. The variability of the thermodilution measurements was greater than that of the lithium dilution measurements. The lithium dilution method is at least as accurate as bolus thermodilution and, since pulmonary artery catheterization is not needed, it has the advantages of being safe and quick to perform.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Linton et al. (1997) studied this question.

synapsesocial.com/papers/6a814a6c2bff219921e3d1a3https://doi.org/10.1097/00003246-199711000-00015
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Assessment of lithium dilution cardiac output as a technique for measurement of cardiac output in dogs2001 · 94 citations
  2. 2Validation of cardiac output measurement with the LiDCO <sup>TM</sup> pulse contour system in patients with impaired left ventricular function after cardiac surgery*2011 · 40 citations
  3. 3Lithium dilution cardiac output measurement: A clinical assessment of central venous and peripheral venous indicator injection*2002 · 66 citations
  4. 4Intra-operative evaluation of a continuous versus intermittent bolus thermodilution technique of cardiac output measurement in cardiac surgical patients1998 · 19 citations
  5. 5Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patients1995 · 146 citations