Key result
Lithium dilution cardiac output measurement shows close agreement with bolus thermodilution within ~0.25 L/min.
Why the study?
This study was conducted to compare the results of cardiac output measurements obtained by lithium dilution and thermodilution.
Does lithium dilution provide accurate cardiac output measurements compared to thermodilution in intensive care and postoperative patients?
Observational (n=40)
No
Does lithium dilution provide accurate cardiac output measurements compared to thermodilution in intensive care and postoperative patients?
Lithium dilution is an accurate alternative to bolus thermodilution for measuring cardiac output, offering the advantage of not requiring pulmonary artery catheterization.
Answers are grounded in this paper and cite their sources.
May support lithium dilution for cardiac output in select ICU cases; hypothesis-generating and requires larger validation before practice change.
Objective To compare the results of cardiac output measurements obtained by lithium dilution and thermodilution. Design Case series, observational study. Setting High-dependency postoperative unit and intensive care unit of a teaching hospital. Patients Forty patients were studied. Thirty-four patients had undergoing heart surgery requiring cardiopulmonary bypass within the previous 2 days; the diagnoses in the other patients were myocardial infarct (n = 2), septicemia (n = 2), adult respiratory distress syndrome, and pericardectomy. Interventions Cardiac output was measured five times in each patient, using lithium dilution (single measurement) and bolus thermodilution (series of three to six measurements according to standard clinical practice, taking the average of the closest three). In a subgroup of 14 patients, cardiac output was also measured using "continuous thermodilution." Measurements and main results Comparing lithium dilution with bolus thermodilution, the mean of the differences (lithium dilution-thermodilution) was -0.25 +/- 0.46 [SD] L/min. Linear regression analysis gave y = 0.31 + 0.89x (r2 = .94) for lithium dilution vs. thermodilution. Conclusions The overall agreement between the two methods was good. The variability of the thermodilution measurements was greater than that of the lithium dilution measurements. The lithium dilution method is at least as accurate as bolus thermodilution and, since pulmonary artery catheterization is not needed, it has the advantages of being safe and quick to perform.
No takes yet. Share an insight, caveat, or question.
Linton et al. (1997) conducted an observational in Postoperative heart surgery or critical illness requiring cardiac output monitoring (n=40). Lithium dilution cardiac output measurement vs. Bolus thermodilution (and continuous thermodilution in 14 patients) was evaluated on Agreement in cardiac output measurements between lithium dilution and bolus thermodilution (MD -0.25 L/min (SD 0.46); r2 = 0.94). Lithium dilution cardiac output measurement showed good agreement with bolus thermodilution.