Comparative analysis shows differences in radiochemical purity measurements for actinium-225 in radiopharmaceuticals using 221fr and 213bi measurements.
Abstract Background: Actinium-225 radiopharmaceuticals have drawn great interest in cancer therapy due to their tumor-specific delivery of cytotoxic alpha-particles. Detection and quality control (QC) are critical for these potent agents. There is currently no consensus for best practice of ²²⁵Ac QC. Detection of ²²⁵Ac (T1/2 = 9.92 days) is challenging; however, the gamma-emitting progenies ²²¹Fr (T1/2 = 4.8 min, 218 keV) and ²¹³Bi (T1/2 = 45 min, 440 keV) facilitate the indirect measurement of ²²⁵Ac. Using several instruments, we compared multiple analytical methods for ²²⁵Ac limit of quantification and the radiopharmaceutical radiochemical purity (RCP%). The RCP% of ²²⁵Ac-radiopharmaceutical was evaluated at both ²²¹Fr and ²¹³Bi secular equilibrium. RCP% was measured using two TLC plate readers: a gas-filled proportional counter, under mixed emission and alpha specific parameters; and a plastic silicon photomultiplier detector and compared. Chromatographic strips were also analyzed using LS, HPGe, and NaI(Tl) gamma well counting. We correlated these RCP% to HPLC results measured at both equilibria utilizing multiple energy windows. Finally, examining accuracy and precision for each instrument, free ²²⁵Ac spiking assays were conducted on the radiopharmaceutical and measured. Results: The TLC plate-reader using the plastic silicon detector resulted in significant RCP% differences between 30 min and >5 hr readings, whereas the gas-filled proportional counter showed no-significant differences between alpha-specific (30 min) and the mixed-isotope setting (>5 hr). HPGe-TLC demonstrated RCP% equivalence at ²²¹Fr and ²¹³Bi equilibrium. NaI(Tl) and LS measurements significantly underestimated the RCP% at 30 min. Finally, gamma well counting of HPLC fractions resulted in 3-5 % RCP% underestimation at ²²¹Fr equilibrium. Conclusion: Five instruments have been tested for their accuracy, sensitivity, linearity and specificity response to ²²⁵Ac quantification using ²²¹Fr and ²¹³Bi measurements. ²²¹Fr secular equilibrium was deemed acceptable for TLC RCP% analyses using HPGe and gas-filled proportional counter under adequate parameters. LS, gamma well counting and TLC-scanner with plastic silicon detector required a reading at ²¹³Bi equilibrium for accurate RCP% characterization. Low content radioimpurity detection was most accurately measured using TLC rather than gamma well counting of HPLC fraction collection due to gamma counting limitations.
No takes yet. Share an insight, caveat, or question.
Toro-González et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: