This research compares 2D-pose and 3D motion capture methods for shoulder ROM, indicating that 2D-pose can provide valid measurements with some limitations.
Background Smartphone-based 2D-pose estimation offers a convenient method for assessing shoulder range-of-motion (ROM), but its accuracy compared to 3D motion capture (3D-mocap) needs to be determined. Methods Shoulder ROM was recorded in seventeen participants without shoulder issues using 3D-mocap and 2D-pose concurrently. Movements included abduction, flexion, extension, external, and functional internal rotation (IR). 2D-pose ROM estimates (mymobility's ® Skeletal Tracking Shoulder Range of Motion Assessments feature (Apple Vision framework, Apple Inc., Cupertino, CA, USA) were compared to 3D-mocap using linear mixed-models and Bland-Altman analysis. The influence of thoracic compensation and anatomical frame definitions on shoulder ROM estimates was examined. Results High consistency was observed between 2D-pose and 3D-mocap ( R 2 > 0.98), especially for abduction and flexion. Differences in ROM were linked to anatomical frame variations and thoracic contributions, with 2D-pose overestimating ROM at greater ranges (2°–25°). Internal rotation zone identification showed high consistency, but 2D-pose-based extension and external rotation showed more variability due to thoracic compensation. Conclusions Smartphone-based 2D-pose estimation provides a valid alternative for shoulder ROM measurement but should not be used interchangeably with 3D-mocap due to discrepancies arising from anatomical frame definitions and thoracic movements. Shoulder ROM assessment requires consideration of these limitations to ensure appropriate clinical interpretation.
No takes yet. Share an insight, caveat, or question.
Hoorn et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: