Answers are grounded in this paper and cite their sources.
This study presents a temperature-tunable liquid lens utilizing silicon oil (SO) to achieve precise focal length modulation without requiring mechanical displacement of the sample, typically required in conventional z-scan methodologies. The lens was fabricated by injecting SO into a cavity formed between two convex glass substrates, with temperature varied from 20 °C to 130 °C to enable a focal length tuning range from 69.30 mm to 77.78 mm, corresponding to an overall change of 8.48 mm. Key performance metrics include a focal length resolution of 7.2×10⁻² mm, temperature sensitivity of 1 °C, and demonstrated long-term stability under repeated thermal cycling. The lens’s functionality was validated within a z-scan setup, where it replaced fixed-focal-length optics to eliminate sample translation along the z-axis. Nonlinear refractive index measurements of silver nanoparticles dispersed in distilled water at a concentration of 13 mM yielded a value of (-1.03 ± 0.08) × 10⁻⁷ cm²/W using the proposed tunable lens, closely matching (-0.97 ± 0.06) × 10⁻⁷ cm²/W obtained via the conventional z-scan method. The proposed approach provides a practical modification to the classical Z-scan technique by allowing real-time focal control, which can simplify experimental procedures and potentially improve measurement accuracy.
No takes yet. Share an insight, caveat, or question.
Kabiri et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: