Study demonstrates effective detection of polycyclic aromatic hydrocarbons in samples, suggesting significant sensitivity enhancements.
This study demonstrates, for the first time, the quantitative detection of polycyclic aromatic hydrocarbons (PAHs) using monoisotopic silver-109 nanoparticle-enhanced targets in laser desorption/ionization mass spectrometry (109AgNPET LDI MS). Ten PAHs — anthracene, benzo(a)anthracene, benzo(a)pyrene, benzo(e)pyrene, chrysene, fluoranthene, perylene, phenanthrene, pyrene, and triphenylene — were analyzed, represented by four distinct molecular formulas, with radical cations (M+•) appearing as the predominant species in the MS spectra. The method yielded low chemical background, limited to silver cluster signals, enabling high-sensitivity analysis without matrix-related interferences. Limits of detection (LOD) ranged from 24 pg/mL for benzo(a)pyrene, benzo(e)pyrene, and perylene to 6.97 ng/mL for benzo(a)anthracene, chrysene, and triphenylene. Compared with existing methods, 109AgNPET LDI MS exhibited sensitivity surpassing or comparable to LC-MS, and in the case of benzo(a)pyrene, equaled the GC-MS, while eliminating sample pre-extraction or concentration steps. The results establish 109AgNPET LDI MS as a viable, rapid, and preparation-free approach for sensitive PAH analysis.
No takes yet. Share an insight, caveat, or question.
Adrian Arendowski (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: