Comparative evaluation of insulin delivery systems reveals differences in glucose homeostasis metrics.
Simultaneous values of glucose rate of change (RoC) and glucose can be presented in a dynamic glucose region plot, and risk spaces can be specified for (RoC, glucose) values expected to remain in the target range (glucose 3.9–10.0 mmol/L) or leave or return to the target range within the next 30 min. We downloaded continuous glucose monitoring (CGM) data for 60 days from persons with type 1 diabetes using two different systems for automated insulin delivery (AID), A (n = 65) or B (n = 85). The relative distribution of (RoC, glucose) values in risk spaces was compared. The fraction of all (RoC, glucose) values anticipated to remain in the target range in the next 30 min was higher with system A (62.5%) than with system B (56.8%) (difference 5.7, 95% CI (2.2–9.2%), p = 0.002). The fraction of (RoC, glucose) values in the target range with a risk of progressing to the above range (glucose > 10.0 mmol/L) was slightly lower in system A than in B (difference −1.1 (95% CI: −1.8–−0.5%, p < 0.001). Dynamic glucose region plots and the concept of risk spaces are novel strategies to obtain insight into glucose homeostasis and to demonstrate clinically relevant differences comparing two AID systems.