Proof of concept study reveals reliable pelvic floor measures in seated proctography, suggesting improved data quality.
We tested the feasibility, data quality, and reliability of an upright magnetic resonance defaecating proctography (uMRDP) technique using an Open 0.5 T ASG MRI scanner. Eight healthy volunteers (2 males) performed seated defaecation on a purpose-built radio-frequency commode coil in an Open scanner. An optimised T2-weighted HASTE sequence captured dynamic changes during all three phases of the Kegel manoeuvre. Inter- and intra-rater variability was measured from the pelvic floor (PF) metrices (M-line, H-line, distances of landmarks) extracted by two radiologists. All relevant PF landmarks could be identified and metrices were extracted with acceptable inter- and intra-rater variability. Intra-rater variation was marginal, with relative absolute differences ranging from 5 to 21% and 3.2-44%. Inter-rater variability was reported using correlation and Bland-Altman plots. Correlation between raters was satisfactory, with r2 > 0.93, and bias ranged from - 1.8 to 0.65 mm. Moreover, the limit of agreement in the Bland-Altman plot ranged from 5.8 to 20.4 mm, indicating satisfactory precision. The proposed uMRDP technique can be used as a feasible and reliable alternative to supine MRDP, without the necessity of gadolinium injection and bowel preparation. It can capture defaecation in a regular seated posture and can provide information complementary to standard-of-care fluoroscopic proctography for clinicians.