Monte Carlo method estimates primary electron beam characteristics in linear accelerators, indicating critical output effects.
To estimate the primary electron beam parameters (PEB), including energy, radial intensity distribution and average angular divergence, of the individual linear accelerator using the Monte Carlo method. A model of the treatment head and a standard field were built by BEAMnrc, and the dose distribution was simulated in water phantoms by DOSXYZnrc to obtain the percentage depth dose curve and off-axis ratio. By debugging the parameters mentioned above until the simulation and measurement results could match. The simulation and measurement results could achieve the best match when the parameters mentioned above were 6.25 MeV, 0.95 mm and 0.1° respectively. The PEB of a linear accelerator could have a significant impact on the output beam characteristics. Monte Carlo estimation is one of the most crucial steps in establishing an individual linear accelerator model.