Direct mapping method and coil-by-coil reconstruction improve image quality in MRI, suggesting new pathways for optimization.
Motivation: Deep learning has been increasingly applied in image reconstruction. However, existing methods have limitations. Goal(s): We propose DL-based multi-coil image reconstruction methods that exploit data redundancy in the coil dimension. Approach: New coil-by-coil reconstruction approaches are proposed, where individual coil images are reconstructed and combined with or without incorporating coil sensitivity. Additionally, a direct mapping model predicts coil-combined images directly from undersampled coil images. Results: Using the proposed methods, high-quality knee images are derived with 6× and 10× acceleration. Incorporation of conventionally calculated coil sensitivity improves coil-by-coil image reconstruction. The direct mapping approach demonstrates slightly better performance even without including coil sensitivity. Impact: Comparing DL-based direct mapping with coil-by-coil reconstruction and incorporating coil sensitivity in different ways are two fundamental questions in multi-coil MRI reconstruction. This work provides evidence that implicit estimation and integrated use of coil sensitivities may provide improved reconstruction.
No takes yet. Share an insight, caveat, or question.
Yan et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: