This research reveals a motion-robust WEPCAST MRI technique that reduces scan duration and improves BBB permeability measurement accuracy.
Motivation: WEPCAST MRI was proposed for non-contrast measurement of BBB permeability, but is limited to long scan duration and sensitive to motion due to its double-subtraction scheme. Goal(s): This study aimed to design a motion-robust and accelerated WEPCAST sequence. Approach: The proposed sequence employed a velocity-selective single-shot acquisition as an alternative to the original phase-contrast for isolating the arterially-labeled spins in the vein. Sequence parameters were optimized. The validity and reproducibility were also examined. Results: VS-WEPCAST reduced scan duration from 6 to 3 minutes, while keeping a strong consistency with the original WEPCAST technique. Test-retest results indicate good reproducibility. Impact: This study proposed an accelerated velocity-selective WEPCAST technique that enables the assessment of global BBB permeability to water in nearly half the original scan time, with improved robustness to motion and cardiac pulsation.
No takes yet. Share an insight, caveat, or question.
Hu et al. (2025) studied this question.