Cryoprobes significantly enhance SNR and inter-hemispheric connectivity in mouse resting-state fMRI, suggesting optimization of neuroimaging techniques.
Motivation: Resting-state functional MRI (rsfMRI) is a valuable neuroimaging technique that is difficult to execute well in mice but can benefit from higher signal-to-noise ratio (SNR). Goal(s): Cryoprobes (CP) provide ~2.5-fold increase in SNR compared to room-temperature (RT) coils. The aim was to determine the added value of a cryoprobe to mouse rsfMRI. Approach: Mice were scanned using the same protocol (anesthesia, scanner, and pulse sequences) but with either a CP or RT coil. The same processing pipeline was used to perform seed-based analysis. Results: Compared to the RT data, the CP data displayed increased inter-hemispheric functional connectivity in the primary somatosensory cortex. Impact: Mouse rsfMRI is difficult but feasible through careful optimization (animal handling, anaesthesia, pulse sequence parameters, etc.). A cryoprobe can significantly improve the quality of mouse rsfMRI studies and may enhance the detection of biological effects with smaller group sizes.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: