Experimental study demonstrates active water cooling of Halbach magnets enhances B0 homogeneity in low-field MRI, indicating potential for improved imaging sequences.
Motivation: Temperature-dependent B0 drift limits the use of sequences (e.g. bSSFP) in low-field MRI, preventing the full utilization of the system's potential. Goal(s): The temperature of the used NdFeB-magnets is stabilized through active water cooling in a Halbach slice, with B0-drift measured and simulated. Approach: A Halbach slice prototype with water-cooled magnets and heating capabilities was designed. The magnetic field was mapped with Hall sensors, while temperature, pressure, and flow sensors characterized the cooling performance. Results: Experimental results confirmed expected homogeneity changes with temperature, while identifying a feasible variant of cooling method for future Halbach MRI systems. Impact: Dynamic temperature-dependent B0-drift of permanent magnet-based low-field MRI represents a significant challenge, especially with bSSFP sequences. Our experimental study, complemented by the construction of a prototype, demonstrates a solution to this problem through active water cooling of the magnets themselves.
No takes yet. Share an insight, caveat, or question.
Povolni et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: